PORTABLE SYNGAS ANALYZER IR-GAS-600P

1

PORTABLE SYNGAS ANALYZER IR-GAS-600P

U

SER

M

ANUAL

WUHAN ENVIRO SOLUTIONS TECHNOLOGY CO.,LTD
Add

:The 4th floor, No.28, Fozuling Road,East-lake

Hi-Tech Development Zone,Industry Park Wuhan
430000,China
Tel86-27-86639266
Mobile

0086-15071131907

Email: [email protected]
Website: www.gasese.com

Содержание IR-GAS-600P

Страница 1: ...IR GAS 600P USER MANUAL WUHAN ENVIRO SOLUTIONS TECHNOLOGY CO LTD Add The 4th floor No 28 Fozuling Road East lake Hi Tech Development Zone Industry Park Wuhan 430000 China Telephone 86 27 86639266 Mobi...

Страница 2: ...e of the IR GAS 600P syngas portable analyser requires a full understanding of the analyzer operation Use the analyzer only as specified in this manual otherwise its good operation and initial perform...

Страница 3: ...service attempts void this warranty Important The sample gas must be free from particles 1 oil traces and its moisture content if not WuHan Tenho can provide portable or on line gas conditioning syste...

Страница 4: ...dance with the following characteristics free of solid particles 1 tar and of oil traces with moisture content between 5 and 95 RH non condensing Before switching ON the analyzer Install the gas analy...

Страница 5: ...ly is switched OFF before connecting a data transmission line to the serial RS232 port Never use the analyser in an EX environment with potential gas explosion risks Notice for analyser application Sm...

Страница 6: ...ways shut off the power supply before servicing the analyzer to prevent from electric shocks In order to operate correctly the analyzer must be maintained timely at least once a year Please consult yo...

Страница 7: ...ical model is as follows most of the organic and inorganic multi atomic and molecular gases in the infrared region have a specific absorption wavelength When the infrared light passes through these ga...

Страница 8: ...01 2 Note Measurement range can be customized by the requirement Max 6 components at the same time Response Time TD T90 10s NDIR Response Time 15min I O port RS232 Work temperature 0 50 Relative humid...

Страница 9: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 9...

Страница 10: ...d be leaded to outdoor Gas outlet pipe should not no blocked 3 Sample gas pressure should be a little bigger than atmosphere pressure 4 Gas flow into the gas analyzer should control at 0 7L 1 2L min r...

Страница 11: ...shows a few second after startup Sampling button press this button to open the pump Record button press this button to record the data The data will be recorded every 10 seconds Max 10 000 groups dat...

Страница 12: ...S 600P 12 3 5 Measurement information In the INFO interface it shows gas sensor information The users cannot modify it only for reference 3 6 Calibration In the calibration interface It needs password...

Страница 13: ...box 5 Function buttons to realize specific functions Zero when doing zero point calibration press zero button Span when doing span point calibration press span button Reset reset to factory calibratio...

Страница 14: ...d as normal d Input the calibration gas value in the blank marked as 4 in the above picture and click SPAN button e g if CH4 is 25 calibration gas then input 25 e Then span calibration is finished 3 R...

Страница 15: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 15 3 7 1 USB connect computer Use USB cable into the USB interface the analyzer connected to the computer and then click on the Connect to PC via USB...

Страница 16: ...sword Note please remember the modified password if you accidentally lost the password you can contact the company after service 3 7 3 Others 3 7 4 Delete data Click on Clear all data click the Confir...

Страница 17: ...covery is successful indicating the successful recovery of factory settings 3 7 6 Touch screen calibration For touch screen calibration if the click speed is too fast may lead to calibration failure T...

Страница 18: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 18 3 7 7 Calorific value setting The settings are completed as shown in the following picture...

Страница 19: ...flow less or no flow solutions Replacing dust filter paper Measurement value larger than the allowable error The instrument is not preheated enough Sampling instability Normal instruments start and p...

Страница 20: ...he use it 3 The instrument should be stored in a clean ventilated and dry environment 4 Strong vibration should be prevented during transportation 6 Contact Information WUHAN ENVIRO SOLUTIONS TECHNOLO...

Страница 21: ...PORTABLE SYNGAS ANALYZER IR GAS 600P 21...