MAC 10
e
3
SERVICE MANUAL
PAGE 7
02/05/03
V. SERVICE : Removal and Installation of the Motor
WARNING !
Disconnect the unit from the electrical power source before attempting any service.
WARNING !
Electrical service should be performed by licensed electricians or authorized ENVIRCO service technicians.
Step 1.
To gain access to the motor/blower, remove the ceiling panel next to the unit, if applicable.
Step 2.
Remove the prefilter off the prefilter frame.(see Figure 5)
Step 3.
Remove the four screws from the electrical cover and tilt the electrical cover open..
Figure 5