Operation Manual
Page 14
4.10. Capacitor
4.10.1. With the back panel removed, remove one screw from capacitor strap.
4.10.2. Slide capacitor out.
4.10.3. Remove the rubber boot and disconnect wires.
4.10.4. Replace with the new capacitor, following the steps above in reverse order.
4.11. Pilot Light
4.11.1 With the back panel removed, disconnect the 3-pin connector and push out
pilot light from the inside of the unit.
4.11.2 Replace the pilot light, following the step above in reverse order.
4.12. Motor
4.12.1. With the back panel removed, disconnect all wire connections between the
motor and other components.
4.12.2. Loosen the Allen Head Set screws on the blower wheel hub and remove the
wheel.
4.12.3. Loosen the screw on the motor belly band until the motor can slide out.
4.12.4. Replace with the new motor, following the steps above in reverse order,
ensuring there is an
⅛
-inch to ¼-inch gap between the blower wheel and the
inlet ring.
CAUTION:
When reinstalling the motor ensure that it does not touch or puncture the
HEPA filter.
4.13. Motor/Blower Lubrication
The IsoClean
unit is equipped with permanently sealed bearings and requires no
lubrication.
4.14. Relocating a Contaminated Unit
NOTE:
Use proper protocol, gowning and protective measures.
4.14.1. If an existing unit is to be placed into storage for an extended period of time,
remove and replace the prefilter as described in Section 4.3.
4.14.2. Turn the unit off and disconnect from power source.
4.14.3. Clean the exterior surfaces of the unit as described in Section 4.2.
4.14.4. Unlock caster wheel locks and move unit to its new location.