STACK 710
6 Theory of Operation and Application
ENVEA
UK Ltd
(PC-000878-MA Issue 7, December 2021)
51
PLCF =
L
e
L
m
where:
L
e
=
the exit pathlength
L
m
=
the measurement pathlength
OPLR =
PLCF
2
Ope
=
1 - 10
-OD x OPLR
Note: The STACK 710 applies the PLCF correction only to the measured opacity. Optical
density is shown as the double-pass value.