MN-38645 • Rev 1 • 04/18 • ivNow Fluid Warmer
Electrical Connection
CAUTION
Power source must match voltage identified
on appliance rating tag. The rating tag
provides essential technical information
required for any appliance installation,
maintenance or repairs. Do not remove,
damage or modify the rating tag.
Grounding reliability can only be achieved when
equipment is connected to an equivalent receptacle
marked “Hospital Grade .”
DANGER
To prevent serious personal injury, death,
or property damage:
Do not use this warmer in the presence of
flammable anesthetic mixtures (with air
or with oxygen or nitrous oxide).
Not category AP or APG equipment
ivNow Power Requirements
120/220/230 V .A .C . — 50/60 Hz, 1 ph
100W per pod (maximum of 400W
with 4 connected pods)
Safety Class I Equipment
No Applied Parts
Mode of Operation: Continuous
NEMA 5-15P
15A - 125V Plug
Hospital Grade
Electrical Information
The power specifi cations are located on the
unit identifi cation nameplate . This nameplate
is permanently attached to the unit and must be
located to verify power requirements .
Preparation
Before operating the module(s), clean the exterior of
the unit with a damp cloth and general hospital cleaner
(isopropyl alcohol) .
Applicable only on units with an equipotential-bonding
terminal:
To prevent an electrical shock hazard between the
appliance and other appliances or metal parts in close
vicinity, an equipotential-bonding terminal is provided .
An equalization bonding lead must be connected to this
stud and the other appliances / metal parts to provide
sufficient protection against potential difference . The
terminal is marked with the following symbol .