Owner’s Manual

MN-38645 • Rev 1 • 04/18

ivNow® 

120V/220V/230V

Содержание ivNow

Страница 1: ...64 N9221 Water St Menomonee Falls WI 53051 Tel 262 251 8356 800 TO B WARM generalinfo enthermics com www enthermics com Printed in the U S A Specifications are subject to change without notice Made in the U S A Owner s Manual MN 38645 Rev 1 04 18 ivNow 120V 220V 230V ...

Страница 2: ...stems The Warming Company www enthermics com MN 38645 Rev 1 04 18 ivNow Fluid Warmer Table of Contents Delivery 1 Transportation and Damage Claims 2 Unpacking 2 Safety Procedures 3 Installation 4 Dimensions 5 Electrical Connection 6 Operation 7 Cleaning and Preventative Maintenance 10 Troubleshooting 11 ...

Страница 3: ...quiries Always include both model and serial number s in any correspondence regarding the appliance Model ______________________________________________ Serial number ______________________________________________ Purchased from ______________________________________________ Date installed ______________________________________________ Environmental Conditions Transport and storage environmental c...

Страница 4: ...livery receipt Driver refuses to allow inspection of containers for visible damage 6 Contact the carrier s office immediately upon finding damage and request an inspection Mail a written confirmation to the carrier s office with the time date and the person called 7 Save any packages and packing material for further inspection by the carrier 8 Promptly file a written claim with the carrier and att...

Страница 5: ...s Any troubleshooting guides component views and parts lists included in this manual are for general reference only and are intended for use by qualified and trained technicians This manual should be considered a permanent part of this appliance This manual and all supplied instructions diagrams schematics parts lists notices and labels must remain with the appliance if the item is sold or moved t...

Страница 6: ...counter top mounted to wall or pole The rear bracket provided with the ivNow can be used to mount the pod s to the wall See mounting instructions on page 5 Note Rear bracket provided with the ivNow is not required for countertop single pod configuration For other mounting options i e Horizontal Rails or roll stands refer to the instructions included with the mounting kit 13 2 335mm 21 2 539mm 8 0 ...

Страница 7: ...top bracket off of the assembly and set aside Remove the rear bracket from the assembly Ensure toggle anchors are fully Fghtened Mount the rear plate to the wall using 1 4 toggle bolts and washers Note For drywall mounFng drywall must be a minimum thickness of 1 2 Use the panel as a template for the hole locaFons Mounting Instructions a b c d ...

Страница 8: ...ower Requirements 120 220 230 V A C 50 60 Hz 1 ph 100W per pod maximum of 400W with 4 connected pods Safety Class I Equipment No Applied Parts Mode of Operation Continuous NEMA 5 15P 15A 125V Plug Hospital Grade Electrical Information The power specifications are located on the unit identification nameplate This nameplate is permanently attached to the unit and must be located to verify power requ...

Страница 9: ...ld at temperature NOTE In the event that fluid should spill inside the cavity unplug the unit to prevent an electrical shock hazard Wipe excess fluid from module immediately Refer to qualified service personnel Qualified service personnel should remove the module control and remove any remaining liquid Perform necessary hospital electrical safety checks before returning the unit to operation ivNow...

Страница 10: ...he control display will be powered The plate will not heat until a bag is placed in module 4 Activate control by placing bag of fluid in module See Figure 1 For optimum performance the bag should be centered over the sensor NOTE The warm up stabilization time will vary slightly depending on the bag temperature and ambient room temperature 5 The display will continue to show the current bag tempera...

Страница 11: ... the STATUS button for 2 seconds to see how long a bag has been at the set temperature The first hour is indicated in minutes and seconds MM SS and subsequent time is indicated in hours and minutes HH MM for the first 24 hours After 24 hours the time is displayed in days and fractions of a day DD DD The timer stops after 30 00 days The time a bag is held at the set temperature will remain in memor...

Страница 12: ...ice personnel Qualified service personnel should remove the module control and remove any remaining liquid Perform necessary hospital electrical safety checks before returning the unit to operation Annual Preventative Maintenance 1 Ensure that the correct Operation and Care Manual is available to all users 2 Ensure that all users have been properly trained in unit s operation 3 Inspect condition o...

Страница 13: ...place the lower control assembly Contact service E B0 PCB Sensor short Contact service E B1 PCB Sensor open E B2 PCB Sensor over temperature E 79 Input voltage high A qualified service technican should check the input voltage to ensure it is at or below 277VAC E 78 Input voltage low A qualified service technican should check the input voltage to ensure it is at or above 85VAC E179 Input voltage op...

Страница 14: ...85 2003 certified company W164 N9221 Water St Menomonee Falls WI 53051 Tel 262 251 8356 800 TO B WARM generalinfo enthermics com www enthermics com Printed in the U S A Specifications are subject to change without notice Made in the U S A ...