MN-28671
Rev 2 • 230V Enthermics Fluid Warming Cabinet Operation & Care Manual • 3
Hazardous
Voltage Present
Medical Equipment classified by Underwriters
Laboratories with Respect to Electrical
Shock, Fire and Mechanical Hazards only, in
Accordance with UL 2601-1 and CAN/CSA
C22.2 No. 601.1.
UL File No.
E201645
Grounding reliability can only be achieved when equipment is
connected to an equivalent receptacle marked “Hospital Grade.”
Protective Earth
Ground Symbol
Safety Class I
Equipment
E L E C T R I C A L I N F O R M A T I O N
The power specifications are located on the unit identification rating tag. This tag is
permanently attached to the unit and must be located to verify power requirements.
P R E P A R A T I O N
Before operating the cabinet, clean both the interior and exterior of the unit with a damp cloth and
mild soap solution. Wipe with an appropriate disinfectant.
Wipe dry with a clean cloth or air dry.
C A U T I O N
THIS UNIT HAS NOT BEEN
APPROVED FOR WARMING OF
BLOOD OR BLOOD PRODUCTS.
CAUTION
INJECTION FLUID MANUFACTURER
SUGGESTS NOT TO WARM INJECTION
FLUIDS OVER 40°C (104°F).
D A N G E R
DO NOT use this warming appliance
in the presence of flammable
anesthetic mixture (with air or
with oxygen or nitrous oxide).
THIS COULD RISK AN EXPLOSION!
(Not category AP or APG equipment )
D A N G E R
ENSURE POWER SOURCE
MATCHES VOLTAGE IDENTIFIED
ON APPLIANCE RATING TAG.
Wire diagram located under top cover of unit
To prevent an electrical shock hazard between the appliance and
other appliances or metal parts in close vicinity, an equalization-
bonding stud is provided. An equalization bonding lead must be
connected to this stud and the other appliances / metal parts
to provide sufficient protection against potential difference.
The terminal is marked with the following symbol.
Other international plugs are available, contact factory for more information.
EC770L
POWER REQUIREMENTS
230 V.A.C. — 50 Hz, 1 ph
0.8 kW, 3.5 Amps
Type B Equipment
BS 1363 Plug
(UK only)
CEE 7/7
220-230V Plug
EC340L
POWER REQUIREMENTS
EC390L
POWER REQUIREMENTS
EC230L
POWER REQUIREMENTS
230 V.A.C. — 50 Hz, 1 ph
0.7 kW, 3.0 Amps
Type B Equipment
BS 1363 Plug
(UK only)
CEE 7/7
220-230V Plug
230 V.A.C. — 50 Hz, 1 ph
0.82 kW, 3.6 Amps
Type B Equipment
BS 1363 Plug
(UK only)
CEE 7/7
220-230V Plug
230 V.A.C. — 50 Hz, 1 ph
0.9 kW, 3.9 Amps
Type B Equipment
BS 1363 Plug
(UK only)
CEE 7/7
220-230V Plug
IMPORTANT
Do not load each basket beyond the recommended
maximum capacity:
EC230L =
16 liters/basket
EC340L =
14 liters/basket (28 total liters/unit)
EC390L =
18 1-liter bottles or 29 1-liter bags
EC770L =
24 liters/basket (72 total liters/unit)
Overloading may cause lower or uneven
temperatures of product and damage to basket
and basket rail supports. Baskets that are
overloaded may slip off rail supports, resulting
in possible damage to product and equipment, as