Page 46 of 52
EN109-1634750-40 IFU
Error messages
When the device is turned on, it will first execute a self-test. When an error is detected, both during
the self-test and during normal operation, a pop-up screen will appear on the display. When the
error is displayed, all outputs will be disabled. When this situation occurs, switch off the unit and
disconnect all accessories. Wait approximately 15 seconds and switch the unit back on.
When the error re-appears, stop using the device and contact the supplier.
Device needs to cool down before you can proceed
The temperature of the device is too high to start treatment. Remove the device from any potential
heat sources (such as direct sunlight) and let the device cool down. Restart the device to continue.
Insufficient or no output current
•
Check the patient cable for breaks or poor contact.
•
Ensure that the sponges are sufficiently damp.
•
Clean the electrodes
•
Check for poor contact
•
In areas where the water is soft, it is possible that the water may conduct poorly resulting in
abnormally low currents. In this case it is possible to use a saline solution to improve the
conductivity of the water.
Water reservoir full
The water separation tank of the Vacuum Module is full. Continue the treatment with the standard
electrodes or empty the tank as follows:
•
Set power line switch [17] Off (0).
•
Detach the hose from the outermost hose nipple [21] and empty the reservoir.
•
Reattach the hose to the hose nipple.
•
Set power line switch [17] On (1).
Vacuum leak
There probably is a leak in the vacuum system. This error is usually preceded by a continuously
running pump trying to reach the set vacuum. To protect the system, the pump is automatically
stopped after a certain amount of time. Inspect the vacuum cables and electrodes, set the vacuum
back to zero and try again. If the failure persists, stop using the device and contact your supplier.