SPECIFICATIONS
TECHNICAL DATA
Mains voltage:
100 - 240 V
Frequency:
50 / 60 Hz
Power consumption:
approx. 40 VA
Load:
20-400 W (5 W increments)
6
–
60 Nm (0.5 Nm increments)
Fuses:
2x T1.6H250V
Dimensions:
190 cm x 177 cm x 55 cm (l x h x w)
Weight:
approx. 80kg
Max. patient weight:
150 kg
Heart rate monitoring:
possible with optional chest belt
Exercise modes:
Levels, Torque, Power, Heart rate
Programs:
12 pre-programmed hill profiles
Parameters:
Time, Speed, Cadence, Distance, Heart rate, Energy
ENVIRONMENTAL CONDITIONS
Operating conditions:
Temperature:
10 - 40 °C
Humidity:
10 - 90%
Air pressure:
700
- 1060 hPa
Storage and transport conditions:
Temperature:
-20
- 70 °C
Humidity:
10 - 90 %
Air pressure:
500 - 1060 hPa
SAFETY AND PERFORMANCE STANDARDS
Medical class:
Class I, Medical Device Regulation (EU) 2017/745
Electrical safety:
Class I type B (according to IEC 60601-1)