5. Calibration
Because of the sensor material performance or the influence of
external factors, there may be errors in a certain range after a
period of time use. Should send the torque gauge to a
specialized testing organization for calibration.
If you have some standard force weights or the other
standard load and some test stand, you may calibrate it also.
① Mount the torque gauge.
② Remove the tare by use of the key
.
③ Enter Calibration interface, as Fig 5
-1.
(Password: 1 2
3)
④
Load a standard force.Now the value in standard
input area is just equal to the current measured value.Wait a
moment for the force stability.
⑤ Press
and
to input the standard force value.
⑥ Press
to enter the next calibration. Press
can interrupt the calibration.
NOTE:
❶
Set the unit of force to the unit used in calibration previously (as shown in 3.2.1 Unit)
❷
Ensure that the tare weight of attachment has been remove before calibration.
❸
You can do any point or points calibration from 1 to 5 point, we recommend 5 points calibration.
Fig 5-1
- 10 -