3.3 Scale
In measurement mode, press         to convert  to the 
other hardness scales or tensile strength (MPa).

Notes:
A)  The conversion value is “---” indicated the conversion is out of range.
B)  When you set the conversion from hardness scale to tensile strength or from 
tensile strength to hardness scale, you must be reset the material group.
C)  Conversion value only supplies the general reference, which may result in 
some offset. Precisely conversion generally needs related comparative tests.

3.2 Impact Direction
In measurement mode, press            button to select the 
impact direction.
The tester can do some correction for test directions 
automatically.

2.4 Browse

The tester can save 9 hardness values, you can browse them after testing.

Press 

 

      button to browse stored data and display the first data  in last 

nine test data, including hardness value HLD, material code, conversion, 
impact direction, date and time etc.

Press      or      button to browse the next or previous data group. 
Press          button back to the measurement mode.

3.4  Clock
The Hardness Tester has built-in real-time 
clock system. 
It must be adjusted when you reinstall the 
batteries or other necessary case. 
Press             to enter the clock adjustment 
mode.

Press      or      to change the value and press          
or          to move item.
Press             to confirm the set and back to measurement mode.

Mat'L

Avg.

HRC

Avg.

HL

M

o

n

th

D

a

y

Year

Hour Minute

Ra   1.6

2  Test 

2.1 Instrument preparation

 Install the battery

 Connect impact device

 Turn on the power

 To confirm the accuracy: Use the standard test block to check the tester 

and make sure it work well. 

2.2 Specimen preparation

Inappropriate specimen will bring the error measured results. It should be 
the necessary processing and preparation according to the original 
conditions of specimen. The preparation of specimen shall be annexed the 
basic requirements for the following:

In the test surface preparation, should be avoided 

cold, heat processing on the specimen surface.

 Test surface should be flat with metallic luster, and 

no oxide skin and other dirt.

The roughness of test surface should be Ra 

≤ 

1.6.

 

The specimen must be of 

sufficient mass and rigidity to 
avoid the displacement or move in 
test.
Generally, the mass of the 
specimen is greater than or equal 
to 5kg can be carried out the test 
directly.
The specimen of 2 ~ 5kg need to 
be clamped to fix, the specimen of 
0.05 ~ 2kg should be coupled . If the mass is lighter than 0.05kg, it is 
unfavorable to Leeb test type D.

 The thickness of specimen shall be not less than 5mm, and thickness 

of homogeneous layer (or surface hardening layer) is not less than 
0.8mm.  

 If the test surface is not flat, the radius of test area should be larger 

than 30mm. And should mount a appropriate support ring.

 The specimen should not be with magnetic.





2.3 Test Procedure

1) Loading

Simply load the impact device by sliding the loading 
tube forward.

2) Locating

Then place and hold the impact device on the surface 
of the test piece at the desired test point. Impact 
indirection should be vertical with the test of surface.

3) Impacting (Measuring)

Trigger the impact by pressing the release button. The 
hardness value will be instantaneously displayed.

4) Read test result

Read out the test result from LCD.

Note: 
Generally, each measurement location of sample is conducted for the five tests. 
The distance between any two impact point should be 

3 mm ; the distance 

between impact position and the edge of the sample should be

3 mm.

Direct                               Clamping                   Coupling
 >5kg                                  2~5kg                       0.05~2kg

Mat'L

Avg.

HRC

Avg.

HL

Hardness value: 786HLD;

The 3rd measurement point;

Mean value: 785HLD

Material Code:"01"(Steel & Casting Steel)

Conversions to HRC: 58.6HRC

Mean value of HRC: 58.5HRC

SCALE

2.5 Print Test Report

The tester can be linked to a wireless printer to print a 
report.
Place the printer near the tester(within 2 m), and turn it 
power on.
Press  

     

for about 2 seconds, a report will be printed.

NOTE:
The original data in memory will be erased automatically when 
the print completed.

                T e s t  R e p o r t

I m p a c t  U n i t  T y p e :   D
M a t e r i a l :  S t e e l  &  C a s t  s t e e l
1   7 8 2   H L D          5 8 . 3  H R C
                  2 0 1 5 - 1 2 - 0 3   1 5 : 2 1
2   7 8 7   H L D          5 8 . 7  H R C
                  2 0 1 5 - 1 2 - 0 3   1 5 : 2 1
3   7 8 6   H L D          5 8 . 6  H R C
                  2 0 1 5 - 1 2 - 0 3   1 5 : 2 2
4   7 8 5   H L D          5 8 . 5  H R C
                  2 0 1 5 - 1 2 - 0 3   1 5 : 2 2
5   7 8 4   H L D          5 8 . 4  H R C
                  2 0 1 5 - 1 2 - 0 3   1 5 : 2 3

S =       5  H L D            0 . 4  H R C
X = 7 8 5   H L D          5 8 . 5   H R C
                 2 0 1 5 - 1 2 - 3   1 5 : 2 5

3  Setup 

3.1 Material Code

In measurement mode, press       button 
to select the material code you will

 

test. 

The material code is listed on the back 
label 

Note:
It is necessary to select the material 
classification. When you did not know the 
material type, you should  reference the 
related material handbook.

MAT

DIR

DIR

HRC

HRB

HB

HV

HSD

MPa

SCALE

SCALE

DIR

MAT

DEL

. 4 .

. 5 .

. 6 .

. 7 .

4  Maintenance   

4.1 Routine maintenance

1)  General considerations
Impact device should avoid collision, the weight stack in the use and 
storage. Put into the case after use. The impact body should is in release 
state when storage.
Should avoid close to the strong magnetic. Any of the components can't 
stick any kinds of oil or lubricant.
2)  Clean the impact body
General every 1~2 thousands of times after test, the impact device tube and 
impact should be cleaned up. 
Method: Rotate off the support ring, release and remove the impact body. 
use the nylon brush to clean up the tube and impact body.
3)  Replacement impact ball
For long-term use, the impact ball may be worn, it should be replaced when 
affect the test precision. 