Status Messages
Installation and Operation Manual - COMTEC
®
6000 ATEX GasEx
66
Doc.-ID: COMGAS_11022020_EN
Error Messages
(cont.)
Error
Message
Relay
contact
O
2
signal
output
Description
O
2
sensor
slope too low
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, test gas flow too low, probe filter damaged,
filter head missing, faulty O
2
sensor.
O
2
sensor
slope too high
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
Incorrect calibration gas, O
2
sensor defective.
O
2
sensor
signal
instable
System
error,
open
The error can only occur only during an O
2
sensor calibration. An error reset can be
carried out by the user. A successful re-calibration also resets the error.
Possible causes:
test gas flow too low, probe filter is damaged, process pressure
fluctuations too high.
COe board
communication
error
20.80mA
unless set
otherwise
The error can occur at any time and signals a failure in communication between the display
circuit board and the COe circuit board. A reset can be carried out by the user. If the error
occurs again, switch the system on and off. If the error persists after restarting the system,
contact an ENOTEC service point.
COe sensor
heater short
circuit
20.80mA
unless set
otherwise
The error can occur at any time and signals a short circuit either by COe sensor heating or
within the connection between the electronic unit and COe sensor. A reset can be carried
out by the user.
Possible causes:
Short-circuit of the COe sensor heater cable at the
terminal points of the electronics or the sensor, short circuit within the probe cable, short
circuit inside the COe sensor
COe sensor
heater short
circuit
20.80mA
unless set
otherwise
The error can occur at any time and signals a break in the circuit of COe sensor heater. A
reset can be carried out by the user, once the fault has been remedied.
Possible causes:
contact problems between COe sensor cable and terminal points of
the electronics respectively the probe, probe cable is damaged, defective COe sensor
Open circuit
COe sensor
heating
20.80mA
unless set
otherwise
The error can occur at any time and signals a break in the circuit of COe sensor. A reset
can be carried out by the user, once the fault has been remedied.
Possible causes:
contact problems COe sensor wire to the terminal points of the electronics and the probe,
probe cable is damaged, defective COe sensor
Open circuit
COe sensor
20.80mA
unless set
otherwise
The error can occur at any time and signals a failure of the electronic components of the
COE board. A reset cannot be carried out by the user. If the error persists after restarting
the system, contact an ENOTEC service point.
COe board
hardware error
A-C
20.80mA
unless set
otherwise
COe sensor calibration failed for one of the following reasons. A reset cannot be carried out
by the user. All entries triggered by this error will then also be deleted.
COe sensor
calibration
failed
20.80mA
unless set
otherwise
COe sensor calibration failed for one of the following reasons. A reset cannot be carried out
by the user. All entries triggered by this error will then also be deleted.
COe sensor
offset too
low/high
20.80mA
unless set
otherwise
The error can only occur only during COe sensor calibration. A reset cannot be
carried out by the user. A successful re-calibration remedies the error. Possible causes:
insufficient supply of reference air.
COe factor
alpha/beta/
gamma invalid
20.80mA
unless set
otherwise
The error can only occur only during COe sensor calibration. It can be reset by the user. A
reset cannot be carried out by the user. A successful re-calibration also remedies the error.
Possible causes:
inadequate reference air supply, process pressure is too high, incorrect
test gas (not test air), Coe sensor defective.
Error ENOTEC
REMOTE
module
System
error, open
Indicates a hardware error of the ENOTEC REMOTE
module.
Possible cause:
the ENOTEC REMOTE
module is defective.
Unlisted errors: Other errors cannot be remedied by the customer.
Please contact an ENOTEC service point
Содержание COMTEC 6000 ATEX GasEx
Страница 10: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 10 Doc ID COMGAS_11022020_EN ...
Страница 11: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 11 ...
Страница 12: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 12 Doc ID COMGAS_11022020_EN ...
Страница 13: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 13 ...
Страница 15: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 15 ...
Страница 16: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 16 Doc ID COMGAS_11022020_EN ...
Страница 17: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 17 ...
Страница 18: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 18 Doc ID COMGAS_11022020_EN ...
Страница 19: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 19 ...
Страница 20: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 20 Doc ID COMGAS_11022020_EN ...
Страница 21: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 21 ...
Страница 22: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 22 Doc ID COMGAS_11022020_EN ...
Страница 23: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 23 ...
Страница 24: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 24 Doc ID COMGAS_11022020_EN ...
Страница 25: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 25 ...
Страница 26: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 26 Doc ID COMGAS_11022020_EN ...
Страница 27: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 27 ...
Страница 28: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 28 Doc ID COMGAS_11022020_EN ...
Страница 29: ...Installation and Operation Manual COMTEC 6000 ATEX GasEx System Description Doc ID COMGAS_11022020_EN 29 ...
Страница 30: ...System Description Installation and Operation Manual COMTEC 6000 ATEX GasEx 30 Doc ID COMGAS_11022020_EN ...