ennoLogic eM870M Insulation Multimeter
Technical Specifications
39
4.2.3 AC Current
Range
Resolution
Accuracy
Voltage Drop
40Hz~400Hz
500
μ
A
0.1μA
±(0.8% +4)
102µV/µA
5000
μ
A
1μA
±(0.8% +4)
50mA
10μA
±(0.8% +4)
1.5mV/mA
500mA
0.1mA
±(0.8% +4)
Notes: above accuracies can be guaranteed within 10% ~ 100% of the full range
4.2.4 DC Current
Range
Resolution
Accuracy
Voltage Drop
500
μ
A
0. 1μA
±(0.2% +2)
102µV/µA
5000
μ
A
1μA
±(0.2% +2)
50mA
10μA
±(0.2% +2)
1.5mV/mA
500mA
0.1mA
±(0.2% +2)
Notes: above accuracies can be guaranteed over the full range
4.2.5 Resistance
Range
Resolution
Accuracy
500
Ω
0.1Ω
±(0.3% +5)
5K
Ω
1Ω
±(0.3% +5)
50K
Ω
10Ω
±(0.3% +5)
500K
Ω
100Ω
±(0.3% +5)
5M
Ω
1KΩ
±(0.3% +5)
50M
Ω
10KΩ
±(0.8% +5)
Notes: above accuracies can be guaranteed over the full range