background image

 

 

 

ENMET

 Corporation 

PO Box 979 

Ann Arbor, MI 48106-0979 

www.enmet.com

  

Manual Part Number 
80003-500 

MCN-13-002, 04/29/13 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

ISA-60M 

with

 

MRI-5175 

O

PERATION AND 

M

AINTENANCE

 

M

ANUAL

 

Andersen Medical Gas
12 Place Lafi

fitte

Madisonville, LA 70447
https://www.TheMedicalGas.com
https://www.Shop.TheMedicalGas.com
1-866-288-3783

Содержание ISA-60M

Страница 1: ...w enmet com Manual Part Number 80003 500 MCN 13 002 04 29 13 ISA 60M with MRI 5175 OPERATION AND MAINTENANCE MANUAL Andersen Medical Gas 12 Place Lafi fitte Madisonville LA 70447 https www TheMedicalG...

Страница 2: ...3 Fault Indications 9 4 4 Alarm Latching 9 4 5 Audio Defeat 9 4 6 Display 10 4 7 Operational Menu 10 5 0 Maintenance 11 5 1 Cleaning Instructions 11 5 2 Maintenance Menu ISA 60M 11 5 2 1 Set 4 20mA T...

Страница 3: ...enu Flow Chart 11 Figure 9 Calibration MRI 5175 13 Figure 9 Calibration MRI 5175 20 Figure 10 MRI 5175 Sensor Replacement 21 List of Tables Table 1 Wiring ISA 60M to MRI 5175 7 Table 2 Relay Failsafe...

Страница 4: ...l not honor any claims for damage This shipment was carefully inspected verified and properly packaged at our company and delivered to the carrier in good condition When it was picked up by the carrie...

Страница 5: ...installed to the instrument Alarm 1 The top center A red alarm LED for all sensor transmitters Alarm 2 Near the center of the panel A green power indicator LED A red fault alarm LED Pushbutton Switche...

Страница 6: ...g Flanges 4 places Menu Select SN XXXX 88 Power Fault XXX XXX XXX O2 by vol O2 O2 Display Front Cover Latch Audio Alarm Fault LED SELECT Switch Power LED OPTION Switch Alarm Defeat Alarm Acknowledge S...

Страница 7: ...Figure 2 Feature Description Relay Terminals This group of terminals is located at the left side of the Circuit Board For the contacts for each of four alarm relays and for the contacts of a fault re...

Страница 8: ...les the ambient air in the MRI environment It is recommended to use 8 drywall anchors and screws for mounting the ISA 60M to a drywall sheetrock surface Non magnetic mounting hardware is required Dime...

Страница 9: ...of the two enclosure punch outs on the side or bottom of the enclosure See Figure 1 Follow state and local guidelines when selecting appropriate strain relief conduit and wiring Remote Sensor Each mus...

Страница 10: ...the circuit board next to the terminal are for the non energized state which is identical to the fault state The PC Board is labeled for the relays in their un energized state If the relay is configur...

Страница 11: ...g of 0 at the bottom of the range 20mA corresponds to a full scale reading of 30 Wiring requirements are the same as for the relays 3 4 4 Digital Outputs The RS232 RS485 connections are used for firmw...

Страница 12: ...The differential Setting has no effect on alarm relays or on the audio alarm for alarm point 2 4 3 Fault Indications Fault indications are associated with sensor zero and calibration activities and ar...

Страница 13: ...taken for a period of 45 seconds it returns to the operational display Figure 7 ISA 60M Operation Menu Flow Chart S S O O O S O Press Option switch S Press Select switch O No Function O No Function S...

Страница 14: ...ation instructions can be found in Section 6 0 5 1 Cleaning Instructions CAUTION Never spray a cleaning solution on the surfaces of the ISA 60M or MRI 5175 devices Clean the exterior of the ISA 60M an...

Страница 15: ...Fault led will show Red for 3 seconds and then flash red green red green red green 6 Press and release the Select switch to exit calibration and return to operation mode If the calibration was complet...

Страница 16: ...e digit to the right when the last digit is accepted the display will read Ch O2 mAXmter NA Trim Press the SELECT switch to enter The display will read O2 20 8 3029 Trim Val 20 7 note this number will...

Страница 17: ...STD is displayed Press the OPTION switch to toggle between STD and DIFF select the correct indicator Press the SELECT switch to lock in the correct indicator IF STD is selected ALARM 1 V L is displaye...

Страница 18: ...d DIFF Locks selection STD S Alarm 1 DIFF BAND 000 v L 19 5 Alarm 1 O S Changes character indicated by underscore cursor Locks underscored character and moves cursor to next digit DIFF S O S Changes c...

Страница 19: ...he OPTION switch to change the left digit select the correct digit Press the SELECT switch which locks the digit in place and moves the cursor one digit to the right Continue this process until the fo...

Страница 20: ...e Set Relay Failsafe Configuration Flow Chart 5 2 10 Set New Password To change the password you must enter the maintenance menu Press the OPTION switch until Enter Maint Menu is displayed then press...

Страница 21: ...ivation Each channel of the ISA 60M can be deactivated and activated as needed CAUTION Deactivated channels will not display or respond to connected sensor transmitters A fault indication occurs when...

Страница 22: ...ibration and then flash red green red green If the calibration was unsuccessful following the 60 to 120 second stabilization time the Power Fault led will show Red for 3 seconds and then flash red gre...

Страница 23: ...for 3 seconds NOTE If the calibration process indicates a failure wait 30 minutes and repeat process or contact ENMET Corporation Figure 9 Calibration MRI 5175 Power Fault LED MENU Switch SELECT Switc...

Страница 24: ...er lid retention screws and remove the lid See Figure 10 Remove the old O2 sensor by unscrewing it and replace it with a new sensor See Figure 10 Replace cover lid with retention screws Turn on the el...

Страница 25: ...Pa Mechanical ISA 60M Dimensions 11 x 9 x 6 inches 4 3 x 3 5 x 2 4 cm Weight 8 lbs 3 6 kg Strain relief 5 12 mm OD Material Engineered thermoplastic with hinged front cover Non magnetic hardware MRI 5...

Страница 26: ...warranty period the customer is responsible for the difference between the most economical means and the expedient mode ENMET shall not be liable for any loss or damage caused by the improper use of...

Страница 27: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 28: ...nt Terms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Ca...

Отзывы: