AM-5150

 

 

ENMET Corporation

 

11 

 
 

N

OTE

:

 Software revision may cause variations of display output. 

Normal Gas Display 

 

 

 

 

 

 

F

IGURE 

7:

 

AM-5150 Maintenance Menu Flow Chart 

5ppm 

M

ENU

  Press and H

OLD

 the M

ENU

 button for 2 – 4 seconds to enter the Maintenance Menus 

Exit 

S

ELECT

 

Press the S

ELECT

 button to return to the Normal Gas Display.  

See Section 5.2.1

 

M

ENU

 

Press the M

ENU

 button to cycle through Maintenance Menus 

OR 

OR 

M

ENU

 

PV: 

 

20 

Apply Cal Gas until signal value becomes stable 
(about 1 to 4 minutes) See Figure 8 

Press S

ELECT

 to enter the cal signal: 

 

If cal is good display will indicate OK or Same 

 

If cal is not within preset “range” display will 

indicate Bad Sens 

 

Cal  OK 

Same  mV 

Bad  Sens 

Span 

S

ELECT

 

S

ELECT

 

S

ELECT

 

You can Press and H

OLD

 the S

ELECT

 button to 

change the Calibration Gas Level 

See Section 5.2.3 

Alarm1 

M

ENU

 

Alarm2 

M

ENU

 

Alarm3 

S

ELECT

 

S

ELECT

 

S

ELECT

 

  20 

   5 

   5 

  10 

  10 

To change Alarm set points: 
Press Menu button until Alarm to be changed is displayed 
Press Select button to display the set point 

The M

ENU

 button changes digit indicated by underscore 

cursor 
The S

ELECT

 button locks underscored digit and moves to 

next digit 

If change is not within range display returns to first digit 
If change is within range display moves to next menu 

See Section 5.2.4 

M

ENU

 

mA Span 

M

ENU

 

S

ELECT

 

0050 

To change mA Span set point: 
Press Menu button until mA Span is displayed 
Press Select button to display the set point 

The M

ENU

 button changes digit indicated by underscore cursor 

The S

ELECT

 button locks underscored digit and moves to next digit 

See Section 5.2.5 

To return to Normal Gas Display: 
Press M

ENU

 button until EXIT is displayed 

Then press S

ELECT

 button 

OR 

Zero 

Cal  OK 

If the Zero signal is not within Preset Specs the AM-5150 will 

display Bad ZERO  

S

ELECT

 

M

ENU

 

Bad ZERO 

If the Zero signal is within Preset Specs the AM-5150 will display 

Cal OK, 

See Section 5.2.2

 

PV: 

S

ELECT

 

Press the S

ELECT

 button to initiate Zero adjustment 

Содержание AM-5150

Страница 1: ...ENMET Corporation PO Box 979 Ann Arbor MI 48106 0979 www enmet com Manual part number 80003 550 MCN 453 07 19 11 AM 5150 Ambient Air Monitor Manual...

Страница 2: ...15 5 3 SENSOR REPLACEMENT 16 5 3 1 Sensor Replacement Calibration 17 6 0 ACCESSORY AND REPLACEMENT PARTS 18 7 0 TECHNICAL DATA AND SPECIFICATIONS 18 8 0 WARRANTY 19 List of Tables TABLE 1 RELAY FAILSA...

Страница 3: ...T it legally became your company s property If your shipment arrives damaged Keep the items packing material and carton As Is Within 5 days of receipt notify the carrier s local office and request imm...

Страница 4: ...ormation is displayed Audio Alarm Horn Audio alarm 105 dB at 30cm 12in The audio alarm is activated when the unit is in alarm Visual Indicators and Alarms LED indicators Power Fault Indicator LED Gree...

Страница 5: ...ircuit Board For the contacts for each of three alarm relays and for the contacts of a fault relay See Section 3 3 Terminal J12 For VDC back up power in and the 4 20 mA output See Section 3 1 2 Sensor...

Страница 6: ...f the installation to appropriate codes is not ENMET s responsibility The AM 5150 should be powered through circuit breakers provided for this purpose 3 1 2 Power Supply The input power can vary from...

Страница 7: ...low Heavier than Air Gas Sensor Location Bottled LP liquefied petroleum Interior wall 18 24 from floor DO NOT locate directly above or beside gas appliances ovens heaters Avoid locating anywhere near...

Страница 8: ...Heater Ground Brown 4 Signal Ground Yellow Wiring for 3 wire Terminal Sensors Position Function Wire Color 1 Heater Orange 2 Signal Blue 3 Heater Ground Brown Wire Nut Signal Ground Yellow NOTE The ye...

Страница 9: ...ly Open J14 K1 Relay 1 NC Normally Closed J14 K1 Relay 1 COM Common Alarm 2 J15 K2 Relay 2 NO Normally Open J15 K2 Relay 2 NC Normally Closed J15 K2 Relay 2 COM Common Alarm 3 J16 K3 Relay 3 NO Normal...

Страница 10: ...r is a flashing C The instrument has failed Calibration The last good calibration values are retained but the sensor may not be responsive to gas A new Calibration should be performed As Soon As Possi...

Страница 11: ...l Scale Press MENU button Display returns to gas measurent Operational Display Flow Chart 4 2 1 Alarm Conditions AM 5150 There are three alarm set points available These alarm points are normally set...

Страница 12: ...ay output TABLE 2 AM 5150 MAINTENANCE MENUS SEQUENCE Example of Display Function Normal Display Mode Measurement of CO Press and hold the MENU button for 2 4 seconds to enter the Maintenance Menu The...

Страница 13: ...0 To change Alarm set points Press Menu button until Alarm to be changed is displayed Press Select button to display the set point The MENU button changes digit indicated by underscore cursor The SELE...

Страница 14: ...5150 instrument before initial calibration overnight is preferred The AM 5150 has been pre calibrated at the factory and initial field calibration should result in only fine tuning to circuit as well...

Страница 15: ...ll be made Cal OK appears on the display and in 1 2 seconds display will change to Span If you wish to Span the sensor press the SELECT button you are now ready to apply gas Proceed to gas span step 2...

Страница 16: ...To exit cal press MENU button until Exit appears and press SELECT button If the sensor is outside of acceptable parameters Bad Span is displayed If the sensor did not respond an incompatible span gas...

Страница 17: ...span range Enter the maintenance menu as shown in Figure 7 AM 5150 Maintenance Menu flow chart 1 Press the MENU button until to display Span is displayed 2 Press the SELECT button to initiate the mA S...

Страница 18: ...s on side of sensor see Figure 9 Example 813 4 the first 3 digits are sensor ID the 4 digit indicates a 4 wire sensor 2 Disconnect the instrument for power 3 Disconnect the orange brown blue and yello...

Страница 19: ...lt in only fine tuning to circuit as well as a way to check that installation is successful It is not necessary to open the enclosure to make adjustment The calibration functions are operated with pus...

Страница 20: ...n calibration gases must be in a background of air do not use gases with an inert gas background such as nitrogen or argon 7 0 Technical Data and Specifications Electrical Power 15 Amp fused branch ci...

Страница 21: ...product The purchaser indemnifies and saves harmless the company with respect to any loss or damages that may arise through the use by the purchaser or others of this equipment This warranty is expre...

Страница 22: ...processing delays ENMET charges a one hour minimum billing for all approved repairs with additional time billed to the closest tenth of an hour All instruments sent to ENMET are subject to a minimum...

Страница 23: ...nt Terms K COD Check one K VISA MasterCard______________________ ________ ________ Card number Expiration Card Code K American Express______________________ ________ ________ Card number Expiration Ca...

Отзывы: