LOW SPEED HANDPIECE

The Drill - Aire Plus is equipped with the Doriot One-piece Handpiece.

                                                                 

Maximum air input, 60 PSI Maximum RPM: 20,000 

Normal operating RPM, 500 – 1500 RPM – (as slow as possible without splattering paste)

Attachments

: Accepts both handpiece burs and Doriot / U-type attachments (prophy angle)

 It is used with a prophy angle to polish the teeth after a scaling procedure. Use only approved 
attachments and polishing compounds

Follow all manufacturers recommendations. 

The water should be turned OFF when using the low speed handpiece (polisher).

No straight handpiece is used with the short-grip Doriot one-piece handpiece. The prophy angle 
attaches directly to the motor.

AIR / WATER SYRINGE

The three-way air / water syringe features

Well balanced design, and smooth styling for comfortable use.

Easy release for a speedy exchange of tips.

Fully autoclavable tips.

This handpiece allows the operator the ability to rinse the operative site with a stream of water or mist
or dry / blow debris with a stream of air. 

Press-in here to release tip

9

Содержание DRILL-AIRE PLUS

Страница 1: ...DRILL AIRE PLUS INTRUCTION MANUAL Engler High Speed Dental Air Unit Rev C 03 18 2021 ...

Страница 2: ...r comfort cots warm water heater circulator with pads and other products We also acquired the Alpha Sonic Ora Sonic and Pro Sonic line of piezo scalers Engler manufactures all of the inserts and tips used in the Engler products as well as many others on the market today in the 18K and 25K frequency range Our repair department has the technical knowledge to repair and maintain a number of dental de...

Страница 3: ...ir water syringe Drill Aire Plus high speed dental air unit high speed drill low speed polisher air water syringe Drill Aire high speed dental air unit high speed drill air water syringe Son Mate II ultrasonic scaler 35 000 RPM handpiece low speed polisher Piezo Mate ultrasonic scaler 35 000 RPM handpiece low speed polisher Tri Mate ultrasonic scaler 35 000 RPM handpiece low speed polisher electro...

Страница 4: ...ify that all parts for the device along with any optional accessories ordered with it are included in this shipment It is imperative that the shipment be inspected immediately upon arrival Should any parts be missing or damaged Engler Engineering must be notified immediately All claims submitted after fifteen days of receipt will not be valid EEC makes every effort to verify that our devices are b...

Страница 5: ...angeable 11 INSTRUCTIONS FOR OPERATION AND MAINTENANCE 12 REGULAR OPERATIONAL CONSIDERATIONS 12 MAINTENANCE KITS 12 START UP AND SHUT DOWN SEQUENCE 13 CLEANING HANDPIECES FOOTSWITCH AND POWER CABLES 13 HIGH SPEED HANDPIECE 14 Cleaning the high speed handpiece 15 HIGH AND LOW SPEED HANDPIECE STERILIZATION PROCEDURES 16 AUTOCLAVE AND CHEMICLAVE ONLY 16 IMPORTANT SAFETY PRECAUTIONS 17 INFECTION CONTR...

Страница 6: ... filter when needed The recommended operating pressure is 80 PSI 5 Connect water to red hose 6 Connect power supply to power plug located at the end of the umbilical bundle 7 Connect power supply to power outlet 8 Test each handpiece To avoid unit damage and or injury confirm input line pressure for the air or nitrogen is set to a maximum of 80 PSI If using a compressor verify your compressor pres...

Страница 7: ...AM Locate the handpieces on the front of your Drill Aire Plus Mini From right to left they are High Speed Handpiece Low Speed Handpiece Air Water Syringe LOW SPEED POLISHER AIR WATER SYRINGE HIGH SPEED DRILL 7 WATER CONTROL KNOB ...

Страница 8: ...lift the handpiece from it s cradle holding the handpiece over a sink turn the power ON by turning the power knob to click on then press the footswitch Water will flow after a few seconds HIGH SPEED HANDPIECE This handpiece is used for advanced dentistry including but not limited to cutting sectioning and shaping cracked or broken teeth repairing preparing cavities etc Flush to remove water Keep w...

Страница 9: ...g compounds Follow all manufacturers recommendations The water should be turned OFF when using the low speed handpiece polisher No straight handpiece is used with the short grip Doriot one piece handpiece The prophy angle attaches directly to the motor AIR WATER SYRINGE The three way air water syringe features Well balanced design and smooth styling for comfortable use Easy release for a speedy ex...

Страница 10: ...WATER ON OFF AIR FLOW ADJUSTMENTS Behind Selector Switch LOW SPEED HIGH SPEED HANDPIECE SELECTOR FOOTSWITCH 10 WATER FLOW CONTROL KNOB ...

Страница 11: ...l cause your handpiece to fail prematurely Verify your handpiece manufacture s specifications before modifying these settings WATER FLOW CONTROL This control adjusts the water flow to the handpieces A water control knob is provided for each handpiece Turn clockwise to decrease flow and counter clockwise to increase flow AIR HANDPIECE PRESSURE GAUGE Gives a visual indication of the air pressure del...

Страница 12: ...cluded with every Drill Aire Plus Mini For the best care and maintenance order the Drill Aire Plus Mini deluxe maintenance kit which includes Lares Handpiece Conditioner from Engler Engineering Corporation optional not included with unit MAINTENANCE KITS There are maintenance kits available for the Drill Aire Plus Mini optional not included with unit These kits are essential to keep your unit work...

Страница 13: ... low and high speed handpieces Prophy angle low and high speed handpieces must be lubricated regularly and after autoclave 3 Disconnect the air nitrogen source 4 Disconnect unit from the water source CLEANING HANDPIECES FOOTSWITCH AND POWER CABLES After each procedure or at least once a day it is suggested that the handpieces and cables be thoroughly cleaned and sterilized Clean the outer surface ...

Страница 14: ...into chuck as far as possible Then fully depress push button hard and fast while simultaneously inserting bur into chuck the rest of the way until fully seated Release push button and insertion is completed Caution Be sure to tug firmly on the bur immediately after completing the insertion procedure to verify full seating and secure retention of the bur before operation 3 To remove a bur fully dep...

Страница 15: ...ce 7 With water toggle switch OFF depress the foot switch pedal for at least 5 seconds to expel excess lube from the handpiece 8 Wipe the handpiece of excess lubricant DAILY HANDPIECE LUBRICATION IS ABSOLUTELY ESSENTIAL Operate the high speed handpiece between 25 32 PSI The handpiece is engineered to attain speeds of 360 000 RPM at 32 PSI Use the brush to remove foreign particles A fine wire is pr...

Страница 16: ...tion to prevent cross contamination Use autoclave bags and pouches with indicators to protect handpiece DON T DON T immerse handpiece in water or chemical disinfectant sterilizer DON T use any type of disinfectant on handpiece DON T sterilize handpiece with bur inserted DON T exceed 275 F 135 C in autoclave or chemiclave DON T dry heat or heat transfer sterilize DON T operate handpiece without bur...

Страница 17: ...r or failure to follow the maintenance schedule described above will automatically void the limited warranty for this product Lares Handpiece Conditioner is available from Engler engineering corporation part 10083 All Lares high speed handpieces may be steam autoclaved or chemiclaved Prior to cycling be sure to have Lares Handpiece Conditioner available with the appropriate nozzle hardware attache...

Страница 18: ...ve or Chemiclave For Extended Unusual Periods Of Time Such As Overnight DO NOT Leave Handpiece Components In Sterilizer After Cycle Is Completed 7 Immediately remove handpiece from autoclave or chemiclave Allow to cool sufficiently prior to handling 18 ...

Страница 19: ...00 max Attachment accepts both handpiece burs and Doriot U type attachments Air Requirements Clean filtered moisture free air with recommended pressure of at least 45 psi Do not exceed 60 psi Operation 2 1 Removing handpiece bur or Doriot attachment from Doriot Handpiece Hold the handpiece in the left hand depress housing ring toward the body of the handpiece while twisting turn to the right to op...

Страница 20: ...External Cleaning Remove angle or attachment from motor Disconnect from hose and clean the external surface thoroughly with a sponge or gauze using warm tap water DO NOT IMMERSE INSTRUMENT Wipe dry with a clean cloth Clean Lubricate Lubricate after every sterilization Using the combination cleaner lubricant available from Engler Engineering Corporation Spray the cleaner lube into the drive air tub...

Страница 21: ...ON T immerse instrument in water or use ultrasonic cleaner DON T use any type of disinfectant chemical or soap on instrument Use of chlorine products aldehydes etc will damage handpiece and void all warranties DON T operate motor while changing speed range changing direction or inserting removing attachments DON T exceed 275o F 29 psi 135o C 2 bar in autoclave DON T dry heat or heat transfer steri...

Страница 22: ...o remove a tip press on the locking collar surrounding the tip socket and pull the used tip straight out of the socket To insert a new tip press locking collar and push tip into the socket as far as it will go Release collar and gently tug on tip before using to ensure that tip is securely locked into socket SYRINGE TIP STERILIZATION 1 Remove contaminated syringe tip 2 Remove all visible signs of ...

Страница 23: ...le IMPORTANT The prophy angle is rated for no more 3 000 R P M High speeds will result in the polished surface heating up spattered polish and shorter prophy angle life Keep the unit set in the prophy range whenever using the prophy angle Always start with a low speed and then adjust to a higher speed as required PROPHY ANGLE INSTRUCTIONS FOR USE Dampen rubber cup Dab a small amount of polishing p...

Страница 24: ... of the head cap and a drop inside the gear cavity 8 Being careful not to cross thread reassemble the prophy angle and wipe off all excess oil Place a new rubber cup onto the head cap and confirm that the gears are meshing properly by rotating the cap it should turn easily If not remove and try again DO NOT use the wrench only finger tighten 9 Slide the prophy angle onto the handpiece and lock the...

Страница 25: ...r to set the direction of rotation to counter clockwise otherwise the Head Cap may fly off when in use Due to differences between Prophy Angle suppliers some Head Caps may unscrew counter clockwise Take care when servicing your Prophy Angle by gently trying clockwise then counter clockwise Too much force may strip the threads on the Prophy Angle or Head Cap which will void the warranty The recomme...

Страница 26: ...rclockwise when facing you to prevent threaded prophy cups from unscrewing during the procedure Not recommended for use above 3000 R P M If the head of the angle becomes hot during use lubricate the rim of the cup screw hole with mineral oil Do not heat over 275 ºF 135 ºC Use only Engler Care Free Prophy Rubber Cups Other brands will not properly seal the angle causing premature wear and voiding t...

Страница 27: ...nd TURNING the red pressure relief valve located near the top of the bottle If bottle is not equipped with a pressure relief valve relieve pressure by unscrewing cap pump assy until a hissing sound can be heard Wait till hissing stops to safely remove cap pump assy Notice Tanks may or may not be supplied with a pressure relief valve Those that do not are designed to bleed off pressure from the pum...

Страница 28: ...top of the pump cap Place 3 5 drops of mineral oil into the oil hole Pump several times to work the oil into the walls of the pump until it moves freely Repeat if necessary Screw the pump assy back into the tank and resume normal operations This process should be repeated as often as necessary or when pumping becomes difficult Black particles found in water bottle indicates that the pump assembly ...

Страница 29: ...ing a few simple steps 1 Every few weeks dispose of water in tank Pour gallon of warm water and 1 to 2 cups of vinegar into the tank and swirl the liquid thoroughly inside the tank The vinegar solution can be left in the tank overnight 2 The vinegar solution can be sent through the unit as well 3 Dispose of the vinegar solution and rinse tank and the unit with clean water thoroughly and completely...

Страница 30: ...60 PSI max to the handpiece High speed handpiece 360 000 RPM max 25 32 PSI max to the handpiece Power supply Input 100 240 V 1 5A 50 60Hz Output 24 VDC 2 5 Amps Control box dimensions 9 X 4 X 6 Net weight 8 lbs Shipping weight 14 lbs Shipping box dimensions 14 X 14 X 14 Engler Engineering Corporation 1099 East 47th Street Hialeah Florida 33013 800 445 8581 305 688 8581 FAX 305 685 7671 Web site ww...