13
CH No.
Description
Setting data
Initial data
030
Output range No. 1 span value
1 - 99999
* Check the output range unit
using CH034 - 037.
See
Specifications
Manual.
031
Output range No. 2 span value
032
Output range No. 3 span value
033
Output range No. 4 span value
034
Output range No. 1 unit
0: Not in use
1: ppm
2: %
1
035
Output range No. 2 unit
1
036
Output range No. 3 unit
2
037
Output range No. 4 unit
2
038
-
-
-
039
-
-
-
040
|
119
Data
for
setting,
checking
by
ENERGY SUPPORT CORP.
120
Zero gas concentration (%)
0.00 - 99.90
See inspection
data
121
Span gas concentration (%)
122
-Span gas concentration (%)
-99.90 - 99.90
123
Air point concentration
0.00 - 99.90
125
to
142
Data
for
setting,
checking
by
ENERGY SUPPORT CORP.
143
Linearization table
Specific values for each sensor
See inspection
data
144
|
179
Data
for
setting,
checking
by
ENERGY SUPPORT CORP.
180
Calibration point selection
5: Zero point
6: Span point
8: Air point
8
181
Calibration start
0: Default value
1: Calibration start
0
182
|
189
Data
for
setting,
checking
by
ENERGY SUPPORT CORP.
190
Heater control mode
0: Heater OFF
1: Constant voltage control
2: Constant resistance control 1
3: Constant resistance control 2
1
(Change not
possible.)
191
Heater voltage setting value (V)
5.00 - 11.00
10.50
(Change not
possible.)
192
Heater
resistance
at
room
temperature (
Ω
)
Specific values for each sensor
See inspection
data
193
Resistance ratio
Specific values for each sensor
See inspection
data
194
to
199
Data
for
setting,
checking
by
ENERGY SUPPORT CORP.