Ener-C CTM Controller 1600 Скачать руководство пользователя страница 7

12

    www.ener-c.com

www.ener-c.com    

13

F. Maintenance

G. Troubleshooting

The device does not 
switch on

Poor cooling 
or heating 
performance

No power

Not connected to the power grid

Little water in the tank

Water on the control panel

Little water in the tank

Incorrect temperature set

Check if the device is plugged in. 
If the mains cable is damaged, disconnect the 
device from the mains and call customer service.

Connect the mains cable to the power source

Fill the water tank and press the ON/OFF button 
to reset the device

Make sure that the control panel is dry

Fill the tank as described in section D. 
CTM Controller® 1600 setup

Set the temperature and press the ENTER button

Ener-C | CTM Controller® 1600 | Operating manual | F. Maintenance / G. Troubleshooting | Rev. 1.1 | 2017-12-12

Ener-C | CTM Controller® 1600 | Operating manual | G. Troubleshooting | Rev. 1.1 | 2017-12-12

If this equipment does cause harmful interference to other devices, which can be determined by turning
the equipment off and on, the user is encouraged to try to correct the interference by one or more of
the following measures:

•  Reorient or relocate the receiving device

•  Increase the separation between the equipment. Connect the equipment into an outlet on a 
  circuit different from that to which the other device(s) are connected

•  Consult the manufacturer or field service technician for help

In the event the device experiences a loss of performance or degradation due to electromagnetic
disturbance, the device is expected to continue to operate safely. The device should not be used less
than 30 cm (12 inches) away from any portable and mobile RF communications equipment.

Message

"---"

"E1"

"E10"

"E11"

"E2"

"E3"

What it means

Water level to low

Water temperature < 4 °C
Equipment off

Water temperature Sensor
Disconnected
Equipment off

Water temperature Sensor
short circuit
Equipment off

Water temperature > 47 °C
Equipment off

Heat sink temperature > 55 °C
Equipment off

What to do

Water refill

Switch the device off and pull out the mains plug. Wait for 15 minutes, switch the 
device back on and enter the desired temperature.
If the Problem occurs again, note the error message, pull out the mains plug and 
call customer service

Contact Ener-C for repair

Contact Ener-C for repair

Switch the device off and pull out the mains plug. Wait for 15 minutes, switch the 
device back on and enter the desired temperature.
If the Problem occurs again, note the error message, pull out the mains plug and 
call customer service

Switch the device off and pull out the mains plug. Wait for 15 minutes, switch the 
device back on and enter the desired temperature.
If the Problem occurs again, note the error message, pull out the mains plug and 
call customer service

Cleaning the control unit

Switch the device off and pull out the mains plug. Use a clean, damp cloth to clean the outside of 
the control unit. Only use mild soap when necessary. Make sure that no water gets into the device. 
The CTM Controller® 1600 may be cleaned in this way as often as you like.

NEVER

 immerse the control unit in water and avoid any water getting into the vent openings.

 

Emptying the control unit

To empty the device, remove the cover of the water tank and tilt the control unit so that the water can 
flow out of the tank opening. Then close the cover again.

NOTE:

 Always empty the control unit prior to transport or storage to prevent accidental leakage.

Storage of the device

If the device is not used for a long period of time, take the tube off the control unit. Let the water out 
of the control unit as described above. Use the water drain key to let any residual water out of the control 
unit, tube and CTM Exchanger®. For complete emptying, the CTM Exchanger® must be inverted so that 
the tube points downwards.
Coil the cable up loosely for storage to avoid damage.

Maintenance

Ener-C recommends to perform an yearly check of the CTM Controller® 1600. Please get in contact 
with Ener-C or your distributor to arrange the maintenance. The equipment needs to be sent back with
the original packaging.

The following section provides tips and instructions in case of difficulties in using the CTM Controller® 1600.
In case the information is not sufficient to solve the issue, unexpected events or changes in the performance 
of the device please contact Ener-C or the authorized distributor in your country.
Contact information are provided at the end of this user manual or on www.ener-c-com.

Poor cooling 
or heating 
performance

Error message 
on the control unit 
display

Kinked tube

Reduced cooling performance at ambient 
temperatures over 25 °C

The device is equipped with several automatic 
switch-off functions to protect the system; 
an error message shows that one of these 
situations has occurred (E1, E10, E11, E2, E3)

Make sure that the tube and CTM Exchanger® 
are not kinked or folded. See section E. Operation

Move the device into a cooler room

Switch the device off and pull out the mains 
plug. Wait for 15 minutes, switch the device 
back on and enter the desired temperature. 
If the problem occurs again, note the error 
message, pull out the mains plug and call 
customer service

Содержание CTM Controller 1600

Страница 1: ...Optimizing Pain Management and Rehabilitation CTM Controller 1600 Operating Manual Important product information Do not throw away or destroy Rév 1 1 2017 12 12 ...

Страница 2: ...o be used for a period of 10 years when subjected to regular maintenance The CTM Exchanger shall be used for one patient only for the extension of the treatment Table of contents A Important safety information 4 B General safety and usage instructions 5 C Components 6 Control unit 7 Control panel 7 D CTM Controller 1600 setup 8 Connecting the tube and the CTM Exchanger 9 Filling the control unit w...

Страница 3: ...t least 30 cm 12 should be left free on all sides of the device in order to ensure sufficient air circulation The CTM Controller 1600 must be filled with water prior to use See CTM Controller 1600 setup for the instructions for setting up and filling the appliance Before filling with water or emptying the device must always be switched off and disconnected from the power supply Never use caustic o...

Страница 4: ...the water tank Air inlet Handle Mains cable The control panel display symbols Low water level Power Timer Heating Cooling FLASHING Enter Temperature scale Timer On Off Warmer Cooler Enter Temperature scale choice Tube CTM Exchanger Control unit Water level indicator on side Ener C CTM Controller 1600 Operating manual C Components Rev 1 1 2017 12 12 Ener C CTM Controller 1600 Operating manual C Com...

Страница 5: ...through the window of the water tank on the side of the device Avoid the tank overflowing 2 3 to 3 4 full is ideal When the tank is full close it again using the cover As soon as the tank has been filled connect the mains cable to a suitable socket and press the ON OFF button to switch the device on The control unit then pumps the water through the tube and the CTM Exchanger which may lead to the ...

Страница 6: ...r to re establish the flow remove the CTM Exchanger from the body and set the temperature to 42 C 108 F for about 7 to 10 minutes The flow of water should then be re established and you can restart the cold therapy WARNING Too long application of cold can lead to frostbites and to local skin damages up to tissue necrosis Recommended procedure for therapeutic heating NOTE always follow the therapy ...

Страница 7: ...customer service Switch the device off and pull out the mains plug Wait for 15 minutes switch the device back on and enter the desired temperature If the Problem occurs again note the error message pull out the mains plug and call customer service Cleaning the control unit Switch the device off and pull out the mains plug Use a clean damp cloth to clean the outside of the control unit Only use mil...

Страница 8: ...ste Electrical and Electronic Equipment WEEE Symbol Product owners should refer to their specific state and government regulatory guide lines for proper disposal NOTE Choose the correct size of the exchanger to completely cover the area to treat with the pad where water circulates Cleaning the CTM Exchanger The CTM Exchanger can be cleaned by wiping the plastic surface with a mild washing up liqui...

Страница 9: ...n an environment as described below Table 2 for Transient Electromagnetic Immunity Immunity Test Electrostatic Discharge ESD EN 61000 4 2 Radiated RF EN 61000 4 3 Electrical Fast transient EN 61000 4 4 Surge EN61000 4 5 Conduct disturbances EN 61000 4 6 Power frequency magnetic field 50Hz 60Hz EN 61000 4 8 Voltage dips short interruptions and voltage variations on power supply input lines IEC 6100...

Страница 10: ...0 V m 80 MHz to 800 MHz E2 10 V m 800 MHz to 2 5 GHz NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflection from structures objects and people a Field strengths from fixed transmitters such as base stations for radio cellular cordless telephones and land mobile radi...

Отзывы: