Operation
Turbimax CUS52D
26
Hauser
•
PE100 and 1.4404/316L
All of the parameters are assigned default values in the firmware and cannot be
changed.
•
Standard customization
It is possible to select the material, surface (matte/shiny) and the internal diameter of
the assembly in which the sensor is installed.
•
Customized advanced
For special adjustments, the following table provides recommendations. Alternatively,
adjustments can be performed by the manufacturer's service department.
Assembly/pipe built-in adapter
Zero adjustment
Upper Limit
Adjustment characteristic
CUA250
0.14
33
1.001
CYA251
0.075
25
1.5
VARIVENT N DN 65
1.28
500
6
VARIVENT N DN 80
0.75
500
6
VARIVENT N DN 100
0.35
500
6
VARIVENT N DN 125
0.20
500
6
1)
Sensor adapter required to fit CUS52D in this assembly,
Application selection
‣
During initial commissioning or calibration at the CM44x, select the appropriate
application for your field of application.
Application Field of application
Unit
Formazine
Drinking water, process water
FNU; NTU; TE/F; EBC; ASBC
Kaolin
Drinking water, filterable matter, industrial water
mg/l; g/l; ppm
PSL
The calibration standard commonly used in Japan for drinking
water turbidity
(dough)
Diatomite
Mineral-based solids (sand)
mg/l; g/l; ppm
1 to 6 points can be calibrated for all applications.
In addition to the factory calibrations, which cannot be changed, the sensor contains 6
additional data records for storing process calibrations or for adjusting them to the
relevant measuring point (application).
1-point and multipoint calibration
1. Before a calibration, rinse the system until all air pockets and fouling have been
removed.
2. In the calibration table, edit the actual values as well as the set points (right and left
columns).
3. Add additional pairs of calibration values, even without measurement in a medium.
When factory calibration data records are duplicated, the value pair 1000/1000 is
automatically generated to map the factory data record 1:1 to the duplicated record.
‣
If a single-point or multipoint calibration is performed following duplication, delete the
value pair (1000/1000) in the calibration table
Lines interpolate between the calibration points.