6 Commissioning
TopClean S CPC 30
84
EHauser
pH calibration
"Manual data entry"
The numeric values for electrode zero point and slope are entered numerically and by
hand.
Calibration pH
"Calibration with manual buffer" / "Calibration with fixed buffer" /
"Calibration with automatic buffer recognition"
Manual buffer
: The buffer pH value is entered manually. The display then shows the
current measured value.
Fixed buffer
: In the Calibration menu from s. page 84 onwards, you can set two buffer
solutions or define them yourself. The selected pH value and buffer type are displayed.
Automatic buffer recognition
: The device automatically recognises the used buffer.
Preselect the buffer types (e.g. E+H) in the Calibration menu.
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CA
Calibration with
data entry
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CAA1
025.0 °C
(–20.0 ... +150.0 °C)
Entry of the temperature
where zero point and slope have been detected.
CAA2
07.00
(–2.00 ... +16.00pH)
IsFET:
act. value
(–500 ...
+500mV)
Entry of the electrode zero point
Confirm by pressing "E"
E2
CAA4
59.16 mV/pH
(5.00 ... 99.00 mV/pH)
Entry of the electrode slope
Confirm by pressing "E"
E2
CAA6
Accept
Reject
Recalibrate
End of calibration
Accept
: Pressing "E" accepts the new calibration data.
Reject
: The data is not accepted, a recalibration won’t
be started.
Recalibrate
: The data is rejected and a new calibration
will be started.
E1
CAA7
Electrode in medium?
Note in display:
If the electrode is back in the medium, so that
measurement can take place?
–
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
CA
Calibration with manual
buffer
(with fixed buffer / automatic
buffer recognition)
Note in display
Display of the type of on-site calibration selected in the
calibration settings.
–
CAB1
025.0 °C
(–20.0 ... +150.0 °C)
Entry of the temperature
where zero point and slope have been detected.
CAB2 / CAB6
025.0 °C
(–20.0 ... +150.0 °C)
Entry of the buffer temperature
(only if "Calibrate with MTC" is selected, see Calibration
menu page 45)
Confirm by pressing "E"
E2
CAB3 / CAB7
Immerse:
pH electrode in buffer 1
Handling instructions
Immerse the electrode in buffer 1 / 2 and then press "E".
–