6 Commissioning
TopClean S CPC 30
44
EHauser
PARAM
⇒
Set up 1
⇒
Calibration
(For redox operating mode)
CODE
CHOICE
(default = bold)
INFO
(E1, 2 = Editor types, s. page 28)
User
settings
J1
On-site calibration type
General settings
Calibration timer
Select calibration menu
On-site calibration type:
Under initial settings for the
function of the CAL key.
General settings:
General calibration settings
Calibration timer:
Clock for calibration
E1
On-site calibration type:
JA1
For redox abs.
Data entry abs.
Calibration abs.
Calibration parameter
Sets the calibration type undertaken when the "CAL"
key is pressed:
Data entry abs
.: Enter the electrode offset in mV.
Calibration abs.
: The electrode offset is calculated from
the difference between the current measured value and
the known buffer value.
E1
For: Redox %:
Data entry abs.
Data entry rel.
Calibration abs.
Calibration rel.
Data entry abs
.: Enter the electrode offset in mV.
Data entry rel.
: Entry of two % calibration points to
which one mV value is assigned.
Calibration abs.
: The electrode offset is calculated from
the difference between the current measured value and
the known buffer value.
Calibration rel.
: Use of a non-toxic and unchanged
sample and buffer.
E1
General settings:
JC3
0120 mV
(1 ... 1500 mV)
Entry of offset deviation
of the mV value for the alarm function
If the offset deviates from the reference offset by the
value entered here, an alarm can be triggered.
E2
Calibration timer:
JD1
Cal timer:
on
Warning:
0001h
___________
Time: 0001:00
Calibration timer
If no calibration is undertaken in the set time, an error
message appears (E115).
Cal timer
: on = activate
Warning
: Enter the time within which a calibration must
take place.
Time
: Display of the remaining time to an error
message.
E2