TopClean S CPC 30
8 Trouble-shooting
Hauser
105
8.2
Response of outputs to errors
8.2.1
Response of current outputs
If an error occurs in the system, an error current is output at the current output. You can
adjust the value of this error current in the Alarm menu (see page 42).
If you have configured the controller for functioning with a current output, no error
current is output on this current output should an error occur.
8.2.2
Response of contacts to errors
You can select the assignment of which instrument error messages trigger an alarm
individually for each error message, (see table on page 98, editing errors on page 42).
In "NAMUR" mode, failure messages (E 001 – E 029) always trigger an alarm.
Behaviour with standard setting
Instrument status
Alarm relay
Limit value /
Controller
Normal operation
Picked-up (fail-safe behaviour)
Appropriate
configuration and
operating status
Alarm
Dropped out
Voltage-free
Dropped out
Dropped out