Technical data
Teqwave T
48
Hauser
Battery
USB port
Power consumption
Power supply failure
The configuration and recorded data are retained in the device memory.
Electrical connection
Cable specification
14.5
Performance characteristics
Max. measured error
Accuracy
Accuracy of concentration measurement
The measuring device can achieve an accuracy of up to 0.01%. The accuracy depends on the
concentration app. The data sheet for the concentration app contains detailed accuracy
information under the "Approximation error" section.
Measuring frequencies
Response time
The response time for displaying the temperature depends on the transfer of heat from the
fluid to the steel.
Influence of variations in
the fluid temperature
If the fluid temperature changes quickly (>1.5 °C/min), the meter displays a error message.
Influence of vibrations
The meter displays a error message as a result of mechanical or acoustic vibrations in the 0.8
to 2.0 MHz range.
Influence of air bubbles
Air bubbles and particles are disturbance factors when measuring with acoustic surface
waves. The diagnostic functions integrated in the measuring device largely prevent an incor-
rect measurement result due to air bubbles or particles.
Battery type
Lithium ion battery
Capacity
2300 mAh
Charging time
Approx. 2.5 h
Battery life
Approx. 8 h, at least 40 measured value recordings
Supply voltage
DC 5 V, 1 A, BCv1.2 standard
Transmitter
Max. 2 W
Temperature
±0.5 K
Speed of sound
(internal measured variable)
±2 m/s
Density
(internal measured variable)
±5 kg/m³
Concentration
Single measurement, value displayed averaged over 10 seconds
Temperature
Single measurement, value displayed averaged over 10 seconds