Stamolys CA71CODcr
Wiring
Hauser
19
The results (measured value+unit of measure) are output in the "Data memory Measured values"
menu.
The calibration results (measured value+unit of measure) are output in the "Data memory-
Calibration factors" menu.
A null modem cable is required (not a crossed one).
The analyzer does not have to be configured for the interface.
The following commands can be sent from the PC, in order to read out data:
• "D" = Data memory-Measured values
• "C" = Data memory-Calibration factors
• "S" = Setup (parameter entry, configuration...)
• "F" = Frequency (current)
4.5
Connection of the sampling vessel
1.
De-energize the analyzer.
2.
Fold out the analyzer frame (
3.
Guide the cable of the level measurement into the analyzer and connect the terminals as
follows:
– Black electric line (BK) to terminal no. 58
– Yellow electric line (YE) to terminal no. 57
– White electric line (WH) to terminal no. 54
– Green electric line (GN) to terminal no. 53
4.
Fold back the analyzer frame.
5.
Energize the analyzer and the stirrer of the sampling vessel.