Wiring
Stamolys CA71CODcr
14
Hauser
4
Wiring
WARNING
!
Device is energized
Improper connection can cause injury or death.
►
The electrical connection must only be carried out by a certified electrician.
►
Technical personnel must have read and understood the instructions in this manual and must
adhere to them.
►
Prior to beginning
any wiring work, make sure voltage is not applied to any of the cables.
4.1
Electrical connection
4.1.1
Quick wiring guide
NOTICE
Reaching the terminal block and avoiding damage
►
You have to fold out the frame of the analyzer to reach the terminal block.
►
After folding the frame into position, check that all the tubes are properly secured and fit
properly into the bottles, as otherwise there is a risk of flooding or incorrect measurement.
Fold out the frame as follows:
1.
Release the two bottom Allen screws (AF 6) by 3-4 rotations (
11, item 1).
2.
Unscrew the two upper Allen screws completely so that the frame folds out. In this way, you
reach the terminal block (item 2).
a0012673
Fig. 11:
Folding out the frame
1
Allen screws AF 6
2
Terminal strip
NOTICE
The device does not have a mains switch
►
Install the device near to a fused socket which is easy to reach.