Wiring
Hauser
17
NOTICE
Terms for special terminals
►
Limit switches 1 and 2 do not need to be connected if the PLC sets its own alarm values at the
analog output.
►
When using a sample conditioning system: Connect terminals 57 and 53 on the analyzer to the
corresponding terminals on the sample conditioning system. For the allocation of these terminals
see the sample conditioning system operating instructions.
►
If there is a 24 V voltage at terminal 57, the analyzer will not begin measurement (sample not
ready). To start measurement, the voltage must stay at 0 V for at least 5 seconds.
►
All external relay contacts have to be connected to 230 V.
4.2
Signal connection
4.2.1
Screening of the analog outputs
The interference suppressor attenuates electromagnetic effects on control, power and signal lines.
After the connection of the data transfer cables clip the interference suppressor (in scope of delivery)
on the cable cores (not on the outer insulation of the cable!). Place the cable screen out of the
interference suppressor and connect it to PE (brass screw with bolt, top right in the connection
compartment) (
14).
Modbus RS485
3)
+
1 +
–
2 –
Sample conditioning
remote control
Input
57
0 V
53
1)
Brass screw with bolt top right in the connection compartment (marked with
*
)
2)
At Modbus module terminals
3)
The Modbus module is already installed and wired up at the factory.
a0001746
Fig. 14:
Interference protection of the signal cable
1
Cable screen (to PE )
2
Signal cable
3
Interference suppressor
4
Cable cores of the signal cable
Function
Designation
Terminal one channel
3
4