StamoLys CA 71 SI
Technical data
Hauser
43
10.4
Performance characteristics
10.5
Environment
10.6
Process
10.7
Mechanical construction
Time between two measurements
t
mes
= reaction time + rinse time + waiting time + rinse again time +
filling time
(min. waiting time = 0 min)
Maximum measured error
2 % of measuring range end
Measuring interval
t
mes
to 120 min
Reaction time
4 minutes
Sample requirement
15 ml (0.004 US.gal.) per measurement
Reagent requirement
3 x 0.2 ml (0.00005 US.gal)
0.86 l (0.227 US.gal) per reagent per month with 10 minute
measuring interval
Calibration interval
0 to 72 h
Rinse interval
0 to 72 h
Rinse time
selectable from 20 to 300 s (standard = 80 s)
Rinse again time
30 s
Filling time
15 s
Maintenance interval
6 month (typical)
Servicing requirement
15 minutes per week (typical)
Ambient temperature
5 ... 40 °C (41... 104 °F), avoid markable fluctuations
Humidity
below the condensation limit, installation in usual, clean rooms
outdoor installation only possible with protective devices
(customer supplies)
Ingress protection
IP 43
Sample flow rate
min. 5 ml (0.0013 US.gal.) per min
Consistance of the sample
low solid content (< 50 ppm)
Sample inlet
pressureless
Design, dimensions
see chapter "Installation"
Weight
GFK housing
approx. 28 kg (61.7 lb)
Stainless steel
housing
approx. 33 kg (72.8 lb)
Materials
Housing
Stainless steel 1.4301 (AISI 304) or GFK
Front windows
Plexiglass
®
Endless hose
C-Flex
®
, Norprene
®
Pump hose
Tygon
®
, Viton
®
Valves
Tygon
®