2-Pack Analyzer System
3
–
8
4900002256 rev. B 4-13-21
4.
Strip back the jacket and/or insulation of the wires just enough to
connect to the power terminal block.
5.
For AC systems, attach the neutral and hot wires to the power
terminal block by
connecting the neutral wire to the terminal marked
“NEU,” the hot wire to the terminal marked “LINE,” as shown in
For DC systems, connect the minus wire to the terminal marked “
,”
and the plus wire to the terminal marked “+,” as shown in Figure 3–3
below.
6.
Connect the ground wire to the ground terminal marked
.
7.
Close and tighten the Analyzer B electronics enclosure door.
Because the breaker in the power distribution panel or switch will
be the primary means of disconnecting the power from the
analyzer, the power distribution panel should be located in close
proximity to the equipment and within easy reach of the operator,
or within 10 feet of the analyzer.
An approved switch or circuit breaker rated for 15 amps should be
used and clearly marked as the disconnecting device for the
analyzer.
Figure 3–3
AC and DC connections for the electronics
control board terminal block in the Analyzer B
electronics enclosure
AC CONNECTIONS
DC CONNECTIONS
Содержание SS1000
Страница 2: ......
Страница 4: ......
Страница 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Страница 121: ...www spectrasensors com contact 4900002256 Rev B ...