Maintenance and Troubleshooting
Hardware Installation and Maintenance Manual
B
–
23
The powder material in the scrubber efficiency indicator changes color from
turquoise to dark grey if there is any H
2
S breakthrough, as shown in Figure
B–20 on page B–25. Alternatively, regular validation of the system with an
appropriate gas standard will indicate when the scrubber needs to be replaced.
The system will activate a
New Scrubber Alarm
fault, which triggers the
General Fault Alarm
to indicate when it is time to replace the scrubber and
scrubber efficiency indicator. Once the scrubber and scrubber efficiency
indicator have been replaced, reset the scrubber lifetime monitor with the
New
Scrub Installed
parameter and the
General Fault Alarm
with the Reset
option for the
General Alarm
DO
parameter (see
“To change parameters in
Mode 2”
in the Firmware Manual for your analyzer).
When specifying gas standards, indicate H
2
S in methane balance.
For a measured range of 0-20 ppm, a concentration of 4 to 16
ppm is recommended.
Figure B–18
Predicted scrubber lifetime based on
average H
2
S load
20
0
40
60
80
100
20
0
40
60
80
100
REMAI
N
IN
G
C
APACI
T
Y
(%
)
0
100
200
300
400
500
600
DAYS
4 ppmv, 2” Scrubber
100 ppmv, 3” Scrubber
Содержание SS1000
Страница 2: ......
Страница 4: ......
Страница 8: ...2 Pack Analyzer System iv 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 12: ...2 Pack Analyzer System viii 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 66: ...2 Pack Analyzer System 4 12 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 84: ...2 Pack Analyzer System A 18 4900002256 rev B 4 13 21 THIS PAGE INTENTIONALLY LEFT BLANK ...
Страница 120: ...2 Pack Analyzer System Index 4 4900002256 rev B 4 13 21 This page intentionally left blank ...
Страница 121: ...www spectrasensors com contact 4900002256 Rev B ...