38
Soliwave FDR16/FQR16
Hauser
Technical data
13�1�4 Transmitting power
• The radiated power is maximum 100 mW e.i.r.p. (equivalent isotrope radiation perfor-
mance).
•
Power density directly in front of the device: Approx. 1 mW/cm
2
•
Power density at a distance of 1 m: Approx. 0.3 µW/cm
2
13.1.5 Antenna opening angle (3 dB)
Approx. ± 12°
13�2
Output
13�2�1 Output signal
Switching output
• 3-wire DC-PNP (positive voltage signal at the switching output of the electronics)
• 2 DC-PNP outputs, antivalent switched
• Max. 200 mA per output, short-circuit proof
•
Switching delay parameterizable (off, 500 ms to 10 s)
13�2�2 Ex connection data
See safety instructions (XA): All data relating to explosion protection are provided
in separate Ex documentation and are available from the Downloads Area of the
Hauser-website. The Ex documentation is supplied as standard with all Ex de
-
vices.
13�3
Power supply
13�3�1 Supply voltage
• U = 18 to 30 V DC
• In accordance with IEC/EN61010 a suitable circuit breaker must be provided for the
measuring device.
•
Voltage source: Non-hazardous contact voltage or Class 2 circuit (North America).
13�3�2 Power consumption
P ≤ 2.4 W
13�3�3 Current consumption
I ≤ 120 mA (without load)
13�3�4 Load
Max. 200 mA
13�4
Environment
13�4�1 Ambient temperature
-20 to +60 °C (-4 to +140 °F)
13�4�2 Storage temperature
See ambient temperature