Graphic Data Manager
Device configuration, application setup
27
2.6.2
Starting/ending batches
Batches are started or ended using the following
• Control input (an individual control input for each batch)
• Operating the device (softkey 3 or under "Main menu -> Batch")
• Preset counter
• Remote access, i.e. using PC operating software or OPC server
• Profibus or Modbus
• Barcode reader
An analysis is saved at the end of a batch. The analysis is displayed under "Main menu - Extras -
Signal analysis".
!
Note!
• The batch status (started or ended) is retained even after a power failure.
• There must be at least 200 ms between 2 of the same batches (stop/start).
• The minimum duration of a batch is 1 s. Batches that are shorter than this are not recorded.
• A new batch cannot be started until the old one is ended.
• The start and end of batches are recorded in the event log.
2.6.3
Batch control using the PC software
Batches can be controlled using the PC software provided under "Extras-> Special unit functions -
> (Select device) -> Remote control":
Fig. 9:
Batch control using PC software
!
Note!
• Not possible if batches are started and stopped via digital input!
• The batch information can only be modified if the batch has stopped. The information entered is
only transferred to the device when the batch is started.
• If the device's user administration is active, a password must be entered to start/end the batch.
The user must have the appropriate rights.