10 Technical data
Prosonic Flow 92
56
EHauser
10.1.5
Power supply
Electrical connection
Potential equalisation
Cable entry
Power supply connection:
• Standard power supply cable connection
Signal cable connection:
• Circular connector, 4 pin
Sensor cable connection:
• BNC connector
Cable specification
Supply voltage
Transmitter:
• Built-in battery
Special type Ni-Cd battery
Continuous operation time up to 5 hours (backlight OFF)
Recharging time 3 hours (power adapter used)
• Special type power adapter
100...240 V AC, 47...63 Hz
Flow measuring sensors:
• Powered by the transmitter
Power consumption
DC: < 12 W (incl. sensors)
Power supply failure
Memory backup with lithium battery (lifetime approx. 5 years)
10.1.6
Performance characteristics
Reference operating
conditions
• Liquid temperature: +28 °C ± 2 K
• Ambient temperature: +22 °C ± 2 K
• Warm-up period: 30 minutes
Installation:
• Inlet run >10 x DN
• Outlet run > 5 x DN
• Sensors grounded.
• The measuring sensors are correctly installed.