5.5.5
Device status / Error messages
You can read the extended device status, in this case, current error messages, via Command “48”.
The command delivers information which are partly coded in bits (see table below).
!
Note!
You can find a detailed explanation of the device status and error messages and their elimination on
Page 80 ff.!
Byte
Bit
Error no.
Short error description (
0
0
001
Serious device error
1
011
Measuring amplifier has faulty EEPROM
2
012
Error when accessing data of the measuring amplifier EEPROM
3
not assigned
–
4
not assigned
–
5
not assigned
–
6
not assigned
–
7
not assigned
–
1
0
not assigned
–
1
not assigned
–
2
not assigned
–
3
not assigned
–
4
not assigned
–
5
not assigned
–
6
not assigned
–
7
not assigned
–
2
0
not assigned
–
1
081
Connection (downwards) sensor/transmitter interrupted
2
not assigned
–
3
not assigned
–
4
084
Connection (upwards) sensor/transmitter interrupted
5
not assigned
–
6
not assigned
–
7
not assigned
–
3
0
not assigned
–
1
not assigned
–
2
not assigned
–
3
111
Totalizer checksum error
4
121
The I/O- and the amplifier board are not compatible.
5
not assigned
–
6
not assigned
–
7
not assigned
–
Содержание Prosonic Flow 90
Страница 4: ...Brief operating instructions Proline Prosonic Flow 90 4 Endress Hauser...
Страница 42: ...4 Wiring Proline Prosonic Flow 90 42 Endress Hauser...
Страница 62: ...5 Operation Proline Prosonic Flow 90 62 Endress Hauser...
Страница 74: ...7 Maintenance Proline Prosonic Flow 90 74 Endress Hauser...
Страница 78: ...8 Accessories Proline Prosonic Flow 90 78 Endress Hauser...
Страница 94: ...9 Trouble shooting Proline Prosonic Flow 90 94 Endress Hauser...
Страница 108: ...10 Technical data Proline Prosonic Flow 90 108 Endress Hauser...
Страница 112: ...Index Proline Prosonic Flow 90 112 Endress Hauser...
Страница 114: ...www endress com worldwide BA068D 06 en 11 04 50099981 FM SGLM6 0...