Proline Promass 84F
6
Hauser
Measuring range in custody
transfer mode
The following are example data for German PTB approval (liquids other than water).
Measuring ranges for liquids in mass flow
Measuring ranges for liquids in volume flow (also LPG)
!
Note!
For information about the other approvals
→
see corresponding certificate.
Operable flow range
Over 20 : 1 for verified device
Input signal
Status input (auxiliary input), HART:
U = 3 to 30 V DC, R
i
= 5 k
Ω
, galvanically isolated.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start.
Status input (auxiliary input), MODBUS RS485:
U = 3 to 30 V DC, R
i
= 3 k
Ω
, galvanically isolated, switch level: ±3 to ±30 V DC, independent of polarity.
Configurable for: totalizer reset, positive zero return, error message reset, zero point adjustment start.
DN
Mass flow (liquids) Q
min
to Q
max
Smallest measured quantity
[mm]
[inch]
[kg/min]
[lbs/min]
[kg]
[lbs]
8
3/8"
1.5 to 30
3.3075 to 66.15
0.5
1.10
15
½"
5 to 100
11.025 to 220.5
2
4.41
25
1"
15 to 300
33.075 to 661.5
5
11.0
40
1½"
35 to 700
77.175 to 1543.5
20
44.1
50
2"
50 to 1000
110.25 to 2205.0
50
110.25
80
3"
150 to 3000
330.75 to 6615.0
100
220.50
100
4"
200 to 4500
441.00 to 9922.5
200
441.00
150
6"
350 to 12 000
771.75 to 26460
500
1102.5
250
10"
1500 to 35000
3307.5 to 77175
1000
2205.0
DN
Volume flow (liquids) Q
min
to Q
max
Smallest measured quantity
[mm]
[inch]
[l/min]
[gal/hr]
[l]
[gal]
8
3/8"
1.5 to 30
23.76 to 475.20
0.5
0.132
15
½"
5 to 100
79.20 to 1584.0
2.0
0.528
25
1"
15 to 300
237.6 to 4752.0
5.0
1.320
40
1½"
35 to 700
554.4 to 11088
20
5.280
50
2"
50 to 1000
792.0 to 15840
50
13.20
80
3"
150 to 3000
2376 to 47520
100
26.40
100
4"
200 to 4500
3168 to 71280
200
52.80
150
6"
350 to 12 000
5544 to 190 080
500
132.0
250
10"
1500 to 35000
23760 to 554 400
1000
264.0