SENSOR DATA
Proline t-mass 65 MODBUS RS485
50
Hauser
MOUNTING SET LENGTH
Prerequisite
This function is
only
available for insertion sensors (t-mass 65I).
Description
Use this function to enter a value for the length of the mounting set (including
the sensor compression fitting).
The unit is taken from the function UNIT LENGTH (
10).
User input
5-digit floating-point number
75 to 900 (mm) or 2.953 to 35.433 (inch) (country dependent
Factory setting
106.0 (mm) or 4.173 (inch) (country dependent
The factory setting value is the length of the G1A compression fitting and
standard Hauser mounting boss.
MODBUS register:
Data type:
Access:
9635
Float
read/write
INSERTION DEPTH
Prerequisite
This function is
only
available for insertion sensors (t-mass 65I).
Description
This function displays the calculated insertion depth for mounting the sensor.
The unit is taken from the function UNIT LENGTH (
10).
For more information on insertion depth calculation, please refer to the
Operating Instructions BA00115D/06.
Display
5-digit floating-point number
MODBUS register:
Data type:
Access:
9637
Float
read
ZERO POINT
Description
This function shows the current zero point correction value for the sensor. The
zero point is determined by the ZERO POINT ADJUST (
38) function.
User input
5-digit floating-point number
–20.000 to +20.000
Factory setting
Depends on calibration.
MODBUS register:
Data type:
Access:
9624
Integer
read/write
FLOW CONDITIONER
Description
Use this function to indicate if the t-mass 65F sensor has been calibrated with or
without a flow conditioner.
Display
0 = WITH
1 = WITHOUT
Factory setting
WITHOUT
MODBUS register:
Data type:
Access:
9626
Integer
read
CALIBRATION DATE
Description
Use this function to display the date of the last factory calibration of the
measuring device.
This date is not updated by the IN-SITU CALIBRATION function.
Options
DD.MM.YYYY
MODBUS register:
Data type:
Access:
9641
String (16)
read/write
Function description, SENSOR DATA group
Содержание Proline t-mass 65 MODBUS RS485
Страница 2: ......
Страница 62: ...www endress com worldwide BA00116D 06 EN 13 10 71123867 FM SGML6 0 ProMoDo...