5.5
"Pressure" Quick Setup menu
Use this Quick Setup to program the individual process pressure for each gas group. If only one gas
group is being used, then only the function PROCESS PRESSURE 1 needs to be programmed,
PROCESS PRESSURE 2 can remain with default settings.
A0009908-en
!
Note!
• The device operates with absolute pressure only. Convert any gauge pressures to absolute
pressure.
• If a pressure compensating input is being used, then the input signal value overrides the manually
programmed value. The pressure input value applies to both gas groups. i.e. 2 independent
pressure values are no longer possible.
• This Quick Setup Gas function is not available if an in-situ calibration function has been
performed on the device as the in-situ calibration curve refers to the sensor power at each
recorded flow point. Therefore, the programmed pressure settings become redundant.
+
+