Installation
Proline Prosonic Flow P
26
Hauser
4. Lock the bracket in place to fix the sensor housing on the sensor holder.
A0043378
14
Fixing the sensor housing
5. Connect the sensor cable to the adapter cable.
This completes the mounting procedure. The sensors can now be connected to the
transmitter via the connecting cables.
• The visible measuring pipe surface must be clean (free from flaking paint and/or rust)
to ensure good acoustic contact.
• If necessary, the holder and sensor housing can be secured with a screw/nut or a lead
seal (not supplied).
• The bracket can only be released using an auxiliary tool (e.g. screwdriver).
Sensor installation – medium/large nominal diameters DN 50 to 4000 (2 to 160")
Installation for measurement via 1 traverse
Requirements
• The installation distance and wire length are known → 16
• Strapping bands are pre-assembled
Содержание Proline Prosonic Flow P
Страница 39: ......
Страница 40: ...www addresses endress com 71508892 71508892...