Nominal diameter and flow rate
The pipe diameter and the flow rate determine the nominal diameter of the sensor. Optimum flow
velocity is between 2 and 3 m/s. The flow velocity (v) must also match the physical properties of
the fluid:
v < 2 m/s: for abrasive fluids such as potter's clay, lime milk, ore slurry etc.
v > 2 m/s: for fluids that tend to cause buildup, such as wastewater sludge etc.
!
Note!
If the flow velocity has to be increased, this can be achieved by reducing the nominal diameter of
sensor (
Flow characteristics of Prosonic Flow C (SI units)
Nominal diameter
Recommended flow rate
Factory setting
[mm]
[inch]
Min./max. full scale value
(v ~ 0.3 or 10 m/s)
Creepage
(v ~ 0.04 m/s)
300
12"
80…2700 m³/h
10 m³/h
350
14"
100…3300 m³/h
15 m³/h
400
16"
130…4400 m³/h
20 m³/h
450
18"
160…5600 m³/h
20 m³/h
500
20"
200…6900 m³/h
30 m³/h
600
24"
300…9900 m³/h
40 m³/h
700
28"
410…13600 m³/h
55 m³/h
–
30"
470…15900 m³/h
65 m³/h
800
32"
540…17900 m³/h
75 m³/h
900
36"
680…22500 m³/h
90 m³/h
1000
40"
850…25000 m³/h
115 m³/h
–
42"
950…27000 m³/h
125 m³/h
1200
48"
1250…30000 m³/h
160 m³/h
–
54"
1550…32000 m³/h
205 m³/h
1400
–
1650…35000 m³/h
220 m³/h
–
60"
1950…37000 m³/h
255 m³/h
1600
–
2200…40000 m³/h
285 m³/h
–
66"
2500…40000 m³/h
305 m³/h
1800
72"
2800…45000 m³/h
360 m³/h
2000
78"
3400…50000 m³/h
450 m³/h