Installation
Proline Prosonic Flow 93 PROFIBUS DP/PA
28
Hauser
3.6.3
Determining installation distances via Applicator
Applicator is a software application for selecting and planning flowmeters. The installation distances
required for installation can be determined without having to commission the transmitter
beforehand.
Applicator is available:
• On a CD-ROM for installation on a local PC
• Via the Internet for direct online entry
→
www.endress.com
→
select country.
On the Internet site, select
→
Instruments
→
Flow
→
Tooling
→
Applicator. In the "Applicator
Sizing Flow" field, select the "Start Applicator Sizing Flow online" link.
Determining installation distances for clamp-on, measuring via one traverse
Determine the installation distances required via Applicator:
• Select the fluid.
• Select the device (e.g. 93P Clamp-on).
• Enter or select measuring point-specific values.
• Select the number of traverses: 1
• Read off the necessary installation distances:
– Wire length: __________
– Sensor distance: __________
Subsequent procedure
The sensors can be installed once the installation distances have been determined:
• Prosonic Flow P (DN 50 to 4000 / 2 to 160")
• Prosonic Flow W
41.
Determining installation distances for clamp-on, measuring via two traverses
Determine the installation distances required via Applicator:
• Select the fluid.
• Select the device (e.g. 93P Clamp-on).
• Enter or select measuring point-specific values.
• Select the number of traverses: 2
• Read off the necessary installation distances:
– Sensor position: __________
– Sensor distance: __________
Subsequent procedure
The sensors can be installed once the installation distances have been determined:
• Prosonic Flow P (DN 15 to 65 / ½ to 2½")
• Prosonic Flow P (DN 50 to 4000 / 2 to 160")
• Prosonic Flow W
43.
Determining installation distances for insertion version, single-path measurement
Determine the installation distances required via Applicator:
• Select the fluid.
• Select the device (e.g. 93W Insert 1Ch).
• Enter or select measuring point-specific values.
• Read off the necessary installation distance:
– Sensor distance: __________
Subsequent procedure
The sensors can be installed once the installation distances have been determined:
• Prosonic Flow W
46.
Содержание Proline Prosonic Flow 93
Страница 2: ......
Страница 168: ...www endress com worldwide BA00076D 06 EN 13 10 71121236 FM SGML6 0 ProMoDo...