Proline Prosonic Flow 91W
Commissioning
Hauser
33
5.4
Group Sensor Setup
Picklist SENSOR SETUP:
• SETUP
• LIQUID
• PIPE DATA
• LINER
• CONFIG. CHANEL
• POS. SENSOR
• QUIT
SETUP: LIQUID
TEMPERATURE
SOUND VEL. LIQUID
PIPE MATERIAL
SOUND VEL.PIPE
CIRCUMFERENCE
PIPE DIAMETER
WALL THICKNESS
LINER MATERIAL
SOUND VEL. LINER
LINER THICKNESS
SENSOR TYPE
SENSOR CONFIG.
CABLE LENGTH
POS.SENSOR/WIRE
LENGTH
SENSOR DISTANCE
LIQUID: LIQUID
TEMPERATURE
SOUND VEL. LIQUID
PIPE DATA: PIPE MATERIAL
SOUND VEL.PIPE
CIRCUMFERENCE
PIPE DIAMETER
WALL THICKNESS
LINER: LINER MATERIAL
SOUND VEL. LINER
LINER THICKNESS
CONFIG. CHANNEL:SENSOR TYPE
SENSOR CONFIG.
CABLE LENGTH
POS.SENSOR:POS.SENSOR/WIRE LENGTH
SENSOR DISTANCE
The following information is required for a successful setup:
• Sound velocity of the liquid
• Operating temperature
• Pipe circumference or pipe outer diameter
• Sound velocity of the pipe material
• Wall thickness
• Sound velocity of the lining material (if present)
• Thickness of the lining (if present)
• Sensor type
• Sensor arrangement (direct or reflection mode)
• Length of the sensor cable
5.5
Troubleshooting
A complete description of all the error messages is provided in the Operating Instructions on the
CD-ROM.
!
Note!
The output signals (e.g. pulse, frequency) of the measuring device must correspond to the
higher-order controller.