Proline Promass Q 300
18
Hauser
Plain text display
With information on cause and remedial measures
Additional information on remote operation → 70
Web server
Plain text display
With information on cause and remedial measures
Light emitting diodes (LED)
Status information
Status indicated by various light emitting diodes
The following information is displayed depending on the device version:
• Supply voltage active
• Data transmission active
• Device alarm/error has occurred
Ex connection data
Safety-related values
Order code for
"Output; input 1"
Output type
Safety-related values
"Output; input 1"
26 (+)
27 (–)
Option
BA
Current output 4 to 20
mA HART
U
nom
= 30 V
U
max
= 250 V
Option
GA
PROFIBUS PA
U
nom
= 32 V
U
max
= 250 V
Option
MA
Modbus RS485
U
nom
= 30 V
U
max
= 250 V
Option
SA
FOUNDATION Fieldbus
U
nom
= 32 V
U
max
= 250 V
Order code for
"Output; input 2";
"Output; input 3"
Output type
Safety-related values
Output; input 2
Output; input 3
24 (+)
25 (–)
22 (+)
23 (–)
Option
B
Current output 4 to 20
mA
U
nom
= 30 V
U
max
= 250 V
Option
D
User configurable input/
output
U
nom
= 30 V
U
max
= 250 V
Option
E
Pulse/frequency/switch
output
U
nom
= 30 V
U
max
= 250 V
Option
F
Double pulse output
U
nom
= 30 V
U
max
= 250 V
Option
H
Relay output
U
nom
= 30 V
I
nom
= 100 mA DC/500 mA AC
U
max
= 250 V
Option
I
Current input 4 to 20 mA U
nom
= 30 V
U
max
= 250 V
Option
J
Status input
U
nom
= 30 V
U
max
= 250 V