Proline Promass Q 300
44
Hauser
Interior cleaning
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Options
• Oil- and grease-free version for wetted parts, without inspection certificate
Order code for "Service", option
HA
• Oil- and grease-free version for wetted parts, with inspection certificate according to British
Standard – BS IEC 60877:1999+ British Oxygen Cleaning – BOC degreasing specifications 00000-
N-S-430-00-01
Order code for "Service", option
HB
Electromagnetic
compatibility (EMC)
As per IEC/EN 61326 and NAMUR Recommendation 21 (NE 21)
For details, refer to the Declaration of Conformity.
Process
Medium temperature range
Standard
–50 to +205 °C (–58 to +401 °F)
Low temperature
–196 to +150 °C (–320 to +302 °F) Order code for "Measuring tube
material", option
LA
T
A
T
F
B
A
A0031121
T
A
Ambient temperature
T
F
Medium temperature
A
Maximum permissible medium temperature at T
A max
= 60 °C (140 °F); higher medium temperatures require a
reduction of the ambient temperature T
A
(derating)
B
Maximum permitted ambient temperature at the maximum specified medium temperature of the sensor
Sensor
Not insulated
Insulated
A
B
A
B
T
A
T
F
T
A
T
F
T
A
T
F
T
A
T
F
Promass Q 300
60 °C (140 °F)
205 °C (401 °F)
–
–
60 °C (140 °F)
110 °C (230 °F)
50 °C (122 °F)
205 °C (401 °F)
Promass Q 300
60 °C (140 °F)
150 °C (302 °F)
–
–
60 °C (140 °F)
110 °C (230 °F)
50 °C (122 °F)
150 °C (302 °F)
1)
Standard version (order code for "Measuring tube mat., wetted surface", option SA)
2)
Standard version (order code for "Measuring tube mat., wetted surface", option SB)
Seals
No internal seals