Proline Promass I 100
6
Hauser
Measuring system
The device consists of a transmitter and a sensor. If a device with Modbus RS485 intrinsically safe is
ordered, the Safety Barrier Promass 100 is part of the scope of supply and must be implemented to
operate the device.
One device version is available: compact version, transmitter and sensor form a mechanical unit.
Transmitter
Promass 100
Device versions and materials:
• Compact, aluminum coated:
Aluminum, AlSi10Mg, coated
• Compact, hygienic, stainless:
Hygienic version, stainless steel 1.4301 (304)
• Ultra compact, hygienic, stainless:
Hygienic version, stainless steel 1.4301 (304)
Configuration:
• Via operating tools (e.g. FieldCare)
• Additionally for device version with onsite display:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with 4-20 mA HART, pulse/frequency/switch
output:
Via Web browser (e.g. Microsoft Internet Explorer)
• Also for device version with EtherNet/IP output:
– Via Web browser (e.g. Microsoft Internet Explorer)
– Via Add-on Profile Level 3 for automation system from Rockwell
Automation
– Via Electronic Data Sheet (EDS)
A0016693
A0016694
A0016695
Sensor
Promass I
• Sensitive fluid handling thanks to straight single-tube system
• Simultaneous measurement of viscosity, flow, volume flow, density and
temperature (multivariable)
• Immune to process influences
• Nominal diameter range: DN 8 to 80 (³⁄₈ to 3")
• Materials:
– Sensor: stainless steel, 1.4301/1.4307 (304L)
– Measuring tubes: titanium Grade 9
– Process connections: stainless steel, 1.4301 (304), wetted parts:
titanium Grade 2
A0016508
Safety Barrier Promass 100
A0016763
• Dual-channel safety barrier for installation in non-hazardous locations
or Zone 2/Div. 2:
– Channel 1: DC 24 V power supply
– Channel 2: Modbus RS485
• In addition to current, voltage and power limitation, it offers galvanic
isolation of circuits for explosion protection.
• Easy DIN rail mounting (DIN 35mm) for installation in control cabinets