Proline Promass E TB2 HART
Commissioning
Hauser
69
Parameters
Description
Selection/
User entry
Factory setting
Reference density unit
Select the unit for reference
density.
Result
The selected unit applies for:
– Current outputs
– Low value partial filled
pipe detection
– High value partial filled
pipe detection
– Simulation process
variable
– Fixed reference density
– Density adjustment (in the
Expert
menu)
Unit picklist
Country-dependent:
• kg/Nl
• lb/scf
Temperature unit
Select the unit for
temperature.
Result
The selected unit applies for:
– Current outputs
– Reference temperature
– Simulation process
variable
Unit picklist
Country-dependent:
•
°
C (Celsius)
•
°
F (Fahrenheit)
Length unit
Select the unit of length for
the nominal diameter.
Unit picklist
Country-dependent:
• mm
• in
Pressure unit
Select the unit for pipe
pressure.
Unit picklist
Country-dependent:
• bar a
• psi a
10.5.3 Carrying out a sensor adjustment
The
Sensor adjustment
submenu contains parameters that pertain to the functionality of the
sensor.
Navigation path
"Setup" menu
®
Advanced setup
®
Sensor adjustment
Structure of the submenu
Sensor adjustment
®
Installation direction
Zero point adjustment control
Parameter overview with brief description
Parameters
Description
Selection/
User entry
Factory setting
Installation direction
Change the sign of the
direction of flow of the fluid.
• Flow in arrow direction
• Flow against arrow
direction
Flow in arrow direction
Zero point adjustment
control
Start the zero point
adjustment.
• Cancel
• Start
Cancel