Proline Promag W 400 HART
Electrical connection
Hauser
41
Remote version
E1
E2
GND
S1
E1
E2
S2
GND
5
7
4 37
42 41
n.c.
n.c.
2
1
1
2
42 41
A
B
6
5
7
8
4 37 36
n.c.
E
E
S
A0032059
14
Remote version terminal assignment
A
Transmitter wall-mount housing
B
Sensor connection housing
1
Electrode cable
2
Coil current cable
n.c. Not connected, insulated cable shields
Terminal No. and cable colors: 6/5 = brown; 7/8 = white; 4 = green; 36/37 = yellow
7.1.4
Shielding and grounding
7.1.5
Requirements for the supply unit
Supply voltage
Transmitter
Order code for "Power supply"
terminal voltage
Frequency range
Option
L
DC 24 V
±25%
–
AC 24 V
±25%
50/60 Hz, ±4 Hz
AC 100 to 240 V
–15 to +10%
50/60 Hz, ±4 Hz
7.1.6
Preparing the measuring device
Carry out the steps in the following order:
1. Mount the sensor and transmitter.
2. Connection housing, sensor: Connect connecting cable.
3. Transmitter: Connect connecting cable.
4. Transmitter: Connect signal cable and cable for supply voltage.
NOTICE
Insufficient sealing of the housing!
Operational reliability of the measuring device could be compromised.
‣
Use suitable cable glands corresponding to the degree of protection.
1. Remove dummy plug if present.