System integration
Proline Promag H 500 PROFIBUS DP
100
Hauser
Assigned compensation values
A compensation value is permanently assigned to the individual Analog Output blocks.
Function block
Compensation value
AO 1
External temperature
AO 2
External density
1)
The compensation values must be transmitted to the device in the SI basic unit
The selection is made via: Expert → Sensor → External compensation
Data structure
Output data of Analog Output
Byte 1
Byte 2
Byte 3
Byte 4
Byte 5
Measured value: floating point number (IEEE 754)
Status
DI module (Discrete Input)
Transmit discrete input values from the measuring device to the PROFIBUS master (Class
1). Discrete input values are used by the measuring device to transmit the state of device
functions to the PROFIBUS master (Class 1).
The DI module cyclically transmits the discrete input value, along with the status, to the
PROFIBUS Master (Class 1). The discrete input value is depicted in the first byte. The
second byte contains standardized status information pertaining to the input value.
Two Discrete Input blocks are available (slot 10 to 11).
Selection: device function
Device function
Factory setting: Status (meaning)
Empty pipe detection
• 0 (device function not active)
• 1 (device function active)
Low flow cut off
Status verification
• Bit 0: Verification status - Check not done
• Bit 1: Verification status - Failed
• Bit 2: Verification status - Busy
• Bit 3: Verification status - Ready
• Bit 4: Verification overall result - Failed
• Bit 5: Verification overall result - Passed
• Bit 6: Verification overall result - Check not done
• Bit 7: Not used
1)
Only available with the Heartbeat Verification application package
Factory setting
Function block
Factory setting
DI 1
Empty pipe detection
DI 2
Low flow cut off