Installation
Proline Promag H 200
14
Hauser
6
Installation
6.1
Installation conditions
No special measures such as supports are necessary. External forces are absorbed by the
construction of the device.
6.1.1
Mounting position
Mounting location
h =
³
2 × DN
h
A0017061
Installation in down pipes
Install a siphon or a vent valve downstream of the sensor in down pipes whose length
h
³
5 m (16.4 ft). This precaution is to avoid low pressure and the consequent risk of damage
to the measuring tube. This measure also prevents the system losing prime, which could cause
air pockets.
h
2
1
A0017064
å
4
Installation in a down pipe
1
Vent valve
2
Pipe siphon
h
Length of down pipe
Содержание Proline Promag H 200