background image

TI047D/24/ae/07.06

Technical Information

Proline Promag 50P, 53P

Electromagnetic Flow Measuring System
Flow rate measurement in chemical or process applications

Application

Electromagnetic flowmeter for bidirectional 
measurement of liquids with a minimum conductivity 
of

≥ 

µ

S/cm:

• Acids and caustic solutions
• Paints
• Pastes,  mashes
• Water, wastewater etc.

• Flow measurement up to 44,000 GPM

(9600 m

3

/h)

• Fluid temperature up to 356°F (180°C)
• Process pressures up to 580 psi (40 bar)
• Fitting lengths to DVGW/ISO

Application-specific lining materials:
• PTFE und PFA

Approvals for hazardous area:
• ATEX, FM, CSA, TIIS

Connection to process control system:
• HART

®

, Profibus

®

 DP/PA, FOUNDATION™

Fieldbus, MODBUS

®

 RS485

Your benefits

Promag measuring devices offer you cost-effective flow 
measurement with a high degree of accuracy for a wide 
range of process conditions.

The 

Proline transmitter concept 

comprises:

• Modular device and operating concept resulting in a 

higher degree of efficiency

• Software options for batching, electrode cleaning and 

for measuring pulsating flow.

• Uniform operating concept

The tried-and-tested 

Promag sensors

 offer:

• No  pressure  loss
• Not sensitive to vibrations
• Simple installation and commissioning

Содержание Proline Promag 50P

Страница 1: ...Application specific lining materials PTFE und PFA Approvals for hazardous area ATEX FM CSA TIIS Connection to process control system HART Profibus DP PA FOUNDATION Fieldbus MODBUS RS485 Your benefit...

Страница 2: ...consists of a transmitter and a sensor Two versions are available Compact version transmitter and sensor form a single mechanical unit Remote version transmitter and sensor are installed separately Tr...

Страница 3: ...ion 0 5 A active 0 4 to 20 mA RL 700 HART RL 250 passive 4 to 20 mA operating voltage VS 18 to 30 V DC Ri 150 Pulse frequency output passive open collector 30 V DC 250 mA galvanically isolated Frequen...

Страница 4: ...x ia 2 to 5000 Hz on off ratio 1 1 pulse width max 10 s Pulse output pulse value and pulse polarity adjustable pulse width configurable 0 05 to 2000 ms PROFIBUS DP interface Transmission technology Ph...

Страница 5: ...input 1 x discrete output 1 x PID Output data volume flow calculated mass flow temperature totalizer 1 to 3 Input data positive zero return ON OFF reset totalizer Link Master LM functionality is supp...

Страница 6: ...with reverse polarity protection Terminal No 27 DP B PA FF MODBUS RS485 B PA FF with reverse polarity protection e Ground terminal for signal cable shield Fieldbus cable RS485 line f Service connector...

Страница 7: ...ent output Ex i active HART 50 T Frequency output Ex i passive Current output Ex i passive HART Ground connection power supply Page 6 Terminal No inputs outputs Order variant 20 21 22 23 24 25 26 27 F...

Страница 8: ...Frequency output Current output HART 53 N Current output Frequency output Status input MODBUS RS485 53 P Current output Frequency output Status input PROFIBUS DP 53 V Relay output 2 Relay output 1 St...

Страница 9: ...ble cross section max 16 AWG 2 5 mm2 A0003194 a signal cable b coil current cable cross section max 16 AWG 2 5 mm2 1 core 2 core insulation 3 core shield 4 core jacket 5 core strengthening 6 cable shi...

Страница 10: ...actor zero point etc Potential equalisation Standard case Perfect measurement is only ensured when the medium and the sensor have the same electrical potential Most Promag sensors have a standard inst...

Страница 11: ...potential by means of the ground terminal provided for the purpose Caution Also observe company internal grounding guidelines Note The ground cable for flange to flange connections can be ordered sep...

Страница 12: ...ion length You can find the dimensions of ground disks on Page 31 Caution Risk of damage from electrochemical corrosion Note the electrochemical insulation rating if the ground disks and measuring ele...

Страница 13: ...ically 5 A Promag 53 Pulse output 0 2 o r 2 mm s o r of reading Current output plus typically 5 A Supply voltage fluctuations have no effect within the specified range F06 5xxxxxxx 05 xx xx xx 000 Max...

Страница 14: ...conse quent risk of damage to the lining of the measuring tube Information on the lining s resistance to partial vac uum can be found on Page 21 It might be necessary to install pulse dampers in syste...

Страница 15: ...It is advisable to install a cleaning valve F06 5xxxxxxx 11 00 00 xx 002 Vertical pipes Install a siphon b or a vent valve a downstream of the sensor in vertical pipes longer than 5 meters This precau...

Страница 16: ...tical orientation This orientation is ideal for self emptying piping systems and for use in conjunction with Empty Pipe Detection F06 5xxxxxxx 11 00 00 xx 004 Horizontal orientation The measuring elec...

Страница 17: ...ere Information on resistance to vibration and shock can be found on Page 20 F06 5xxxxxxx 11 00 00 xx 006 Foundations supports If the nominal diameter is 14 DN 350 mount the transmitter on a foundatio...

Страница 18: ...r pipes The resultant increase in the rate of flow improves measuring accuracy with very slow moving fluids The nomogram shown here can be used to calculate the pressure loss caused by reducers and ex...

Страница 19: ...0 C Caution It is not allowed to use the device beyond the min and max lining specified temperature values Medium temperature range Note the following points Install the device at a shady location Avo...

Страница 20: ...1 to 8 DN 25 to 200 for restrictions refer to diagrams Compact version PFA and PTFE lining TA ambient temperature TF fluid temperature HT high temperature version with insulation m temperature range f...

Страница 21: ...measuring tube lining Limit values for abs pressure psi mbar at various fluid temperatures mm inch 75 F 25 C 175 F 80 C 212 F 100 C 265 F 130 C 300 F 150 C 355 F 180 C 15 1 2 PTFE 0 0 0 1 45 100 25 1...

Страница 22: ...min 0 20 dm3 0 5 dm3 min 25 1 9 300 dm3 min 75 dm3 min 0 50 dm3 1 dm3 min 32 1 1 4 15 500 dm3 min 125 dm3 min 1 00 dm3 2 dm3 min 40 1 1 2 25 700 dm3 min 200 dm3 min 1 50 dm3 3 dm3 min 50 2 35 1100 dm...

Страница 23: ...0 50 gal 0 75 gal min 2 50 10 to 300 gal min 75 gal min 0 50 gal 1 25 gal min 2 1 2 65 16 to 500 gal min 130 gal min 1 gal 2 0 gal min 3 80 24 to 800 gal min 200 gal min 2 gal 2 5 gal min 4 100 40 to...

Страница 24: ...Cl 150 20K 0 91 23 1 02 26 32 PN 40 20K 1 26 32 1 38 35 40 1 1 2 PN 40 Cl 150 20K 1 42 36 1 61 41 50 2 PN 40 Table E Cl 14 Cl 150 10K 1 89 48 2 05 52 65 PN 16 10K 2 48 63 2 64 67 80 3 PN 16 Cl 150 10K...

Страница 25: ...c E 8 46 215 1 77 45 3 54 90 5 31 135 9 84 250 6 27 159 5 3 56 90 5 2 08 53 2 08 53 3 74 95 7 55 192 0 45 11 5 0 45 11 5 3 20 81 5 4 01 102 3 18 81 3 18 81 1 97 50 13 9 355 6 57 167 10 4 265 9 45 240...

Страница 26: ...ng It can be ordered from Endress Hauser as an accessory The following installation variants are possible Panel mounted installation Pipe mounting Panel mounted installation A0001131 Pipe mounting A00...

Страница 27: ...3 7 94 50 2 7 87 200 13 3 341 10 1 257 3 31 84 4 72 120 3 7 94 65 7 87 200 15 4 391 11 1 282 4 29 109 7 08 180 3 7 94 80 3 7 87 200 15 4 391 11 1 282 4 29 109 7 08 180 3 7 94 100 4 9 84 250 15 4 391 1...

Страница 28: ...20 3 7 94 50 2 7 87 200 11 2 286 7 95 202 3 31 84 4 72 120 3 7 94 65 7 87 200 13 2 336 8 94 227 4 29 109 7 08 180 3 7 94 80 3 7 87 200 13 2 336 8 94 227 4 29 109 7 08 180 3 7 94 100 4 9 84 250 13 2 33...

Страница 29: ...22 2 564 10 9 276 400 16 23 6 600 31 1 790 5 18 9 482 5 12 1 308 0 24 2 616 10 9 276 450 18 25 6 650 33 1 840 5 19 9 507 5 13 1 333 0 26 2 666 11 5 292 500 20 25 6 650 35 1 891 5 20 9 533 0 14 1 358 5...

Страница 30: ...1 1 282 0 22 2 564 10 9 276 400 16 23 6 600 29 0 735 5 16 8 427 5 12 1 308 0 24 2 616 10 9 276 450 18 25 6 650 30 9 785 5 17 8 452 5 13 1 333 0 26 2 666 11 5 292 500 20 25 6 650 32 9 836 5 18 8 478 0...

Страница 31: ...3 3 15 80 5 16 131 6 08 154 5 5 31 135 100 4 4 09 104 6 14 156 7 34 186 5 6 02 153 125 130 187 206 5 160 150 6 6 22 158 8 54 217 10 1 256 7 24 184 200 8 8 11 206 10 5 267 11 3 288 8 07 205 250 10 10...

Страница 32: ...Weight data in kg Nominal diameter Compact version Remote version without cable Sensor Wall housing mm inch EN DIN AS JIS ANSI EN DIN AS JIS ANSI 15 1 2 PN 40 6 5 10K 6 5 Class 150 14 lbs PN 40 4 5 10...

Страница 33: ...ference curves for various process connections relating to the fluid temperature But the maximal permissible fluid temperature always depends on the lining material of the sensor and or the sealing ma...

Страница 34: ...terial A105 RSt37 2 S235JRG2 St44 2 S275JR F06 xxFxxxxx 05 xx xx xx 010 Class 300 Class 150 0 10 20 30 40 50 60 bar 0 145 290 435 580 725 870 psi 76 60 40 40 4 20 32 0 68 20 104 40 140 60 176 80 212 1...

Страница 35: ...ariables Totalizer Promag 50 2 totalizers 7 digit plus 7 digit overflow with sign and units Promag 53 3 totalizers 7 digit plus 7 digit overflow with sign and units Operating elements Unified operatio...

Страница 36: ...ed devices of other manufacturers interoperability FOUNDATION Fieldbus certification The flow device has successfully passed all the test procedures carried out and is certified and registered by the...

Страница 37: ...Promag 53 PROFIBUS DP PA BA053D 06 en BA054D 06 en Operating Instructions Promag 53 FOUNDATION Fieldbus BA051D 06 en BA052D 06 en Operating Instructions Promag 53 MODBUS BA117D 06 en und BA118D 06 en...

Страница 38: ...ompact and remote sensor sizes up to 12 only S NEMA 4X IP 67 remote wall mounted housing HE harsh environment remote sensor sizes 14 to 24 only for approvals A and R available as TSP only T NEMA 4X IP...

Страница 39: ...only up to 12 sensor P NEMA 4X IP 67 compact aluminum housing HE harsh environment compact and remote sensor sizes up to 12 only S NEMA 4X IP 67 remote wall mounted housing HE harsh environment remote...

Страница 40: ...43 Tel 317 535 7138 Sales 888 ENDRESS Service 800 642 8737 fax 317 535 8498 inquiry us endress com www us endress com TI047D 24 ae 07 06 2006 Endress Hauser Canada Endress Hauser Canada 1075 Sutton Dr...