Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
85
Density source
Navigation
Expert → Sensor → External comp. → Density source (6615)
Description
Use this function to select the density source.
Selection
• Fixed density
• Current input 1
• Current input 2
• Current input 3
• External density
• Calculated value
Factory setting
Fixed density
External density
Navigation
Expert → Sensor → External comp. → External density (6630)
Prerequisite
The
External density
option is selected in the
Density source
parameter (→ 85).
Description
Use this function to enter the density read in from the external device.
User entry
Positive floating-point number
Factory setting
0 kg/l
Additional information
Dependency
The unit is taken from the
Density unit
Fixed density
Navigation
Expert → Sensor → External comp. → Fixed density (6623)
Prerequisite
The
Fixed density
option is selected in the
Density source
Description
Use this function to enter a fixed value for the density.
User entry
Positive floating-point number
Factory setting
Country-specific:
• 1
000 kg/l
• 1
000 lb/ft³
*
Visibility depends on order options or device settings