Proline Promag 300 Modbus RS485
Description of device parameters
Hauser
65
Additional information
Result
The defined unit is shown as an option in the choose list of the following parameters:
•
Mass flow unit
•
Mass unit
Example
If the text GLAS is entered, the following options are displayed in the picklist for the
Mass
flow unit
parameter (→ 58):
• GLAS/s
• GLAS/min
• GLAS/h
• GLAS/d
Mass offset
Navigation
Expert → Sensor → System units → User-spec. units → Mass offset (0562)
Description
Use this function to enter the offset for adapting the user-specific mass unit and mass flow
unit (without time).
User entry
Signed floating-point number
Factory setting
0
Additional information
Description
Value in user-specific unit = (factor × value in base unit) + offset
Mass factor
Navigation
Expert → Sensor → System units → User-spec. units → Mass factor (0561)
Description
Use this function to enter a quantity factor (without time) for the user-specific mass and
mass flow unit.
User entry
Signed floating-point number
Factory setting
1.0
Corr. vol. text
Navigation
Expert → Sensor → System units → User-spec. units → Corr. vol. text (0592)
Description
Use this function to enter a text for the user-specific unit of the corrected volume and
corrected volume flow. The corresponding time units (s, min, h, d) for mass flow are
generated automatically.